Teledyne BDS-3960 operating instructions Span Cal

Page 50

Maintenance

BDS 3960

 

 

 

 

 

 

A zero calibration is not recommended during the first 10 days of the operation of the cell.

The ZERO button on the Main screen is used to enter the zero calibration function. Zero calibration can be performed either automatically or manually.

In the automatic mode, an internal timer will purge sensor for five minutes before registering the zero offset of the sensor plus electronics. As the timer counts down, you can force the analyzer to accept the current zero offset using the Accept button. Pressing the Accept button anytime during the countdown period forces the analyzer to accept the zero offset calibration.

Pressing Cancel will return the analyzer back to the analyze mode without accepting the zero calibration. In this case, the previous zero calibration values will remain as the current values.

4.4.1.1ZERO FAILURE

The analyzer checks the output of the cell at the end of the span. If the raw output of the cell produces an Analog to Digital converter count less than 4000 on amplifier gain 0, the zero will not be accepted. The analyzer will use the previous calibration values, and display at the top of the main screen:

ZERO-CALIBRATION

ERROR

4.4.2Span Cal

The SPAN button on the main screen is used to span calibrate the analyzer. Span calibration can be performed automatically or manually. In the automatic mode, an internal timer will purge sensor for five minutes before taking in the zero offset of the sensor plus electronics. Again, the ACCEPT button can be pressed at anytime during the countdown period to accept the zero offset calibration.

Pressing the CANCEL button will return the analyzer to the analyze mode without accepting the zero calibration. The previous zero calibration will still be in effect.

Teledyne Analytical Instruments

50

Image 50
Contents Oxygen Analyzer Teledyne Analytical InstrumentsBDS Copyright 2002 Teledyne Analytical InstrumentsSpecific Model Information IiiSafety Messages Oxygen Analyzer Table of Contents Operation ViiIndex Teledyne Analytical Instruments MaintenanceAppendix ViiiList of Figures BDS-3960 Front PanelVacuum Degassing for the BDS Oxygen Sensor List of Tables Combustible GAS Usage XiiOverview Main Features of the AnalyzerIntroduction Typical ApplicationsFront Panel Operator Interface BDS-3960 Front Panel Remote Probe Rear Panel Equipment InterfaceAnalog Outputs Range ID ContactsModel BDS 3960 Rear Panel BDS Sensor Operational TheoryIntroduction Principles of OperationOxygen Analyzer Operational Theory Cross Section of the BDS Oxygen Sensor Gas Flow Rate BDS sensor output at different gas flow rateGas Pressure Temperature effectRecovery from High Level Oxygen Exposure Background Gas Compatibility Typical Purge-down Curve After Air SaturationMaintenance StabilitySample System Electronics and Signal Processing Flow DiagramOxygen Analyzer Operational Theory BDS 3960 Electronics Block Diagram Page Mounting the Analyzer InstallationUnpacking the Analyzer Oxygen Analyzer Installation Model BDS 3960 Front PanelRear Panel Connections Gas ConnectionsElectrical Connections Equipment Interface Connector Pin Arrangement Threshold Alarm Pin FunctionSystem Alarm Pin ContactCal Contact Installation Electrolyte Refill of BDS Sensor Testing the System Adding Electrolyte to the BDS SensorPowering Up the System Operation Auto Ranging onOxygen Analyzer Maintenance Analyzer applicationSystem Screen Communication Information and Calibration ParametersSetting Software Parameters Filter, Gas Factor, Tmp. Coeff Gas Factor Parameter Selection BoxTemperature Coefficient Parameter Default Values Calibration of the Analyzer Zero CalSpan Cal Alarms Function Maintenance Setting the Analog Output Ranges Range FunctionRange Options List Box Signal Output Fixed Range AnalysisSwitching The Program Back To Front Oxygen Analyzer Maintenance Maintenance Routine Maintenance Adding Water to the BDS SensorFuse Replacement Adding Water into the BDS sensorBattery Backup Replacement Removing Fuse Block from HousingReinstalling Application software to PPC Click startProgramsexplorer Click showMy DeviceSymptoms Possible causes and Solutions CleaningTroubleshooting Major Internal ComponentsCauses and solution keys Vacuum Degassing for the BDS Oxygen Sensor Oxygen Analyzer Maintenance Teledyne Analytical Instruments Specifications AppendixOperating Temperature 5-40 C Oxygen Analyzer AppendixRecommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountApplication notes Section II Physical and Chemical Data Material Safety Data SheetSection I Product Identification Section III -Physical Hazards Section IV Health Hazard DataSection VI Handling Information Alarms IndexOxygen Analyzer Index MsdsZero