Maintenance | BDS 3960 | |
|
|
|
|
|
|
60 | 0.6 | 13.6 |
70 | 0.7 | 15.2 |
80 | 0.8 | 16.8 |
90 | 0.9 | 18.4 |
100 | 1.0 | 20.0 |
The analog output signal has a voltage which depends on the oxygen concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument is currently on, especially when the analyzer is in the autoranging mode.
To provide an indication of the range, a second pair of analog output terminals are used. They generate a steady preset voltage (or current when using the current outputs) to represent a particular range. Table
Table
Range | Voltage (V) | Current (mA) |
|
|
|
Range 1 | 0.25 | 8 |
|
|
|
Range 2 | 0.50 | 12 |
|
|
|
Range 3 | 0.75 | 16 |
|
|
|
Range 4 | 1.00 | 20 |
|
|
|
4.9Switching The Program Back To The
Front
There are times that the Analyzer program may drop out of sight from the front of the screen even though it is still running. If Analyzer program icon is tapped to run the program, an error message will be
Teledyne Analytical Instruments | 56 |