1-2 SYSTEM OVERVIEW

a.Scope

This Instruction Bulletin is designed to sup- ply details needed to install, start up, oper- ate, and maintain the Rosemount Two-Wire In Situ Oxygen Analyzer. The analyzer con- sists of an oxygen probe and Model 3081 Transmitter. The signal conditioning elec- tronics of the Model 3081 Transmitter out- puts a 4-20 mA signal representing an O2 value. An infrared remote control (IRC) al- lows access to setup, calibration, and diag- nostics. This same information, plus additional details, can be accessed with the HART Model 275 handheld communicator or Asset Management Solutions (AMS) software.

b.System Description

The Rosemount Two-Wire In Situ Oxygen Analyzer is designed to measure the net concentration of oxygen in an industrial pro- cess; i.e., the oxygen remaining after all fu- els have been oxidized. The oxygen probe is permanently positioned within an exhaust duct or stack and performs its task without the use of a sampling system. The Model 3081 Transmitter is mounted remotely and conditions the oxygen probe outputs.

The equipment measures oxygen percent- age by reading the voltage developed across an electrochemical cell, which con- sists of a small yttria-stabilized, zirconia disc. Both sides of the disc are coated with porous metal electrodes. The millivolt output voltage of the cell is given by the following Nernst equation:

EMF = KT log10(P1/P2) + C

Where:

NOTE

For best results, use clean, dry, in- strument air (20.95% oxygen) as the reference air.

NOTE

The probe uses a Type B thermocou- ple to measure the cell temperature.

When the cell is at 550°C to 1600°C (1022°F to 2912°F) and there are unequal oxygen concentrations across the cell, oxy- gen ions will travel from the high oxygen partial pressure side to the low oxygen par- tial pressure side of the cell. The resulting logarithmic output voltage is approximately 50 mV per decade.

The output is proportional to the inverse logarithm of the oxygen concentration. Therefore, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the Rosemount Two-Wire In Situ Oxygen Ana- lyzer to provide exceptional sensitivity and accuracy at low oxygen concentrations.

Oxygen analyzer equipment measures net oxygen concentration in the presence of all the products of combustion, including water vapor. Therefore, it may be considered an analysis on a "wet" basis. In comparison with older methods, such as the portable apparatus, which provides an analysis on a "dry" gas basis, the "wet" analysis will, in general, indicate a lower percentage of oxy- gen. The difference will be proportional to the water content of the sampled gas stream.

c. System Configuration

1.P2 is the partial pressure of the oxygen in the measured gas on one side of the cell.

2.P1 is the partial pressure of the oxygen in the reference air on the opposite side of the cell.

3.T is the absolute temperature.

4.C is the cell constant.

5.K is an arithmetic constant.

The equipment discussed in this manual consists of two major components: the oxy- gen probe and the Model 3081 Transmitter.

Oxygen probes are available in three length options, providing in situ penetration appro- priate to the size of the stack or duct. The options on length are 20 in. (508 mm), 26 in. (660 mm), or 34.625 in. (880 mm).

Rosemount Analytical Inc. A Division of Emerson Process Management

Description and Specifications 1-3

Image 13
Contents Model 3081FG Essential Instructions Highlights of Changes Page Model 3081FG Table of ContentsProper Calibration Check Gas Flow Rate List of IllustrationsModel 3081 Transmitter Terminal Block List of TablesModel 3081FG Definitions PrefaceModel 3081FG Component Checklist of Typical System Package Contents Section Description and SpecificationsCode Mounting Adapter Stack Side Product MatrixSystem Overview System ConfigurationScope System DescriptionSystem Features Handling the Analyzer System Considerations Transmitter Typical System InstallationProbe SpecificationsOhms Lift OFF Maximum Power Supply VoltageVDC Operating Load RegionPRE-INSTALLATION Section InstallationMechanical Installation InspectSide View Bottom View Front ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 3081FG Installing Model 3081 Transmitter Circuit Terminal END Cover LockBlock TB Mounting HolesElectrical Installation Display Positioning AssemblyTerminal Block Conduit Model 3081 Transmitter 4-20 mA and Signal ConnectionsThermocouple CELL-WH CELL+BK Reference Air Package Pneumatic InstallationInstrument Air Reference Air Calibration Check GasOutlet Outlet ConnectionInlet Connection Flow SET MAX Point Knob Mounting HolesSection Startup and Operation Power UPGeneral Proper Calibration Check Gas Flow Rate Reestablishing Proper Calibration Check GAS Flow RateOverview OperationDisplay Menu TreeCalcheck DiagnosticsMenu ProgramNavigation Infrared Remote Control IRCAnalyzer Access Code O2 Concentration Analyzer Access Code Or SecurityProgram Menu O2 ConcentrationO2 Concentration Upper Cell Temperature Fault Value O2 Concentration Indicator Fault ValueUpper Range VAL Cell T HITo Reset Maximum Cell Temperature Reset MAX Cell TO2 Concentration Analyzer Access Code SET O2 Filter TimeSET LO Bottle O2 SET HI Bottle O2Concentration Model 3081 Transmitter Parameters Parameter Range DefaultModel 3081 Transmitter Parameters SET Code Show Fault Diagnostics MenuC mV 24. Previous Slope Previous Constant 23. Cell ImpedanceSW Build Date SW Build NumberCalcheck Menu Accept LOW O2 27. in MANUAL?29. Accept LOW O2 Slope Model 3081FG Section HART/AMS Hart Communicator Signal LineOverview ConnectionsMethod 2, For Load Resistance 250 Ohms Analog Output Device Loop ConnectorsHart Communicator PC TWO-WIRE in Situ Oxygen AnalyzerOFF-LINE and ON-LINE Operations Menu Tree for Hart CommunicatorDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 3081FG Model 3081 Electronics Replacement Section Maintenance and ServiceDisplay Board Replacement PC Board Stack ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Oxygen Probe Replacement Oxygen Probe Terminal Block Probe Life Section TroubleshootingIMPEDANCE, Ω Ofresponse Fault IndicationsSpeed Thermocouple Identifying and Correcting Fault IndicationsFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 6, Cell Impedance Too High Fault 5, O2 Cell Open11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 CellModel 3081FG Section Return of Material Model 3081FG Replacement Parts List Replacement PartsSection Index No Part Number DescriptionModel 3081FG Section Index Model 3081FG Model 3081FG Page Section Drawings and Schematics Model 3081FG Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no