Emerson 3081FG instruction manual Essential Instructions

Page 2

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!

Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.

Read all instructions prior to installing, operating, and servicing the product.

If you do not understand any of the instructions, contact your Rosemount Analytical repre- sentative for clarification.

Follow all warnings, cautions, and instructions marked on and supplied with the product.

Inform and educate your personnel in the proper installation, operation, and mainte- nance of the product.

Install your equipment as specified in the Installation Instructions of the appropriate In- struction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.

To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.

When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per- formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.

Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.

The information contained in this document is subject to change without notice.

Emerson Process Management

Rosemount Analytical Inc.

Process Analytic Division

1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010

F (330) 684-4434

e-mail: gas.csc@EmersonProcess.com http://www.processanalytic.com

Image 2
Contents Model 3081FG Essential Instructions Highlights of Changes Page Table of Contents Model 3081FGList of Illustrations Proper Calibration Check Gas Flow RateList of Tables Model 3081 Transmitter Terminal BlockModel 3081FG Preface DefinitionsModel 3081FG Section Description and Specifications Component Checklist of Typical System Package ContentsProduct Matrix Code Mounting Adapter Stack SideScope System ConfigurationSystem Overview System DescriptionSystem Features Handling the Analyzer System Considerations Typical System Installation TransmitterSpecifications ProbeVDC Lift OFF Maximum Power Supply VoltageOhms Operating Load RegionMechanical Installation Section InstallationPRE-INSTALLATION InspectFront View Side View Bottom ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 3081FG Installing Model 3081 Transmitter Block TB Cover LockCircuit Terminal END Mounting HolesDisplay Positioning Assembly Electrical InstallationThermocouple CELL-WH CELL+BK Model 3081 Transmitter 4-20 mA and Signal ConnectionsTerminal Block Conduit Instrument Air Reference Air Pneumatic InstallationReference Air Package Calibration Check GasInlet Connection Outlet ConnectionOutlet Flow SET MAX Point Knob Mounting HolesGeneral Power UPSection Startup and Operation Reestablishing Proper Calibration Check GAS Flow Rate Proper Calibration Check Gas Flow RateDisplay OperationOverview Menu TreeMenu DiagnosticsCalcheck ProgramInfrared Remote Control IRC NavigationProgram Menu O2 Concentration Analyzer Access Code Or SecurityAnalyzer Access Code O2 ConcentrationUpper Range VAL O2 Concentration Indicator Fault ValueO2 Concentration Upper Cell Temperature Fault Value Cell T HIO2 Concentration Analyzer Access Code Reset MAX Cell TTo Reset Maximum Cell Temperature SET O2 Filter TimeConcentration SET HI Bottle O2SET LO Bottle O2 Model 3081 Transmitter Parameters SET Code Parameter Range DefaultModel 3081 Transmitter Parameters C mV Diagnostics MenuShow Fault 23. Cell Impedance 24. Previous Slope Previous ConstantCalcheck Menu SW Build NumberSW Build Date 27. in MANUAL? Accept LOW O229. Accept LOW O2 Slope Model 3081FG Overview Hart Communicator Signal LineSection HART/AMS ConnectionsAnalog Output Device Loop Connectors Method 2, For Load Resistance 250 OhmsOFF-LINE and ON-LINE Operations TWO-WIRE in Situ Oxygen AnalyzerHart Communicator PC Menu Tree for Hart CommunicatorDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 3081FG Display Board Replacement Section Maintenance and ServiceModel 3081 Electronics Replacement PC Board Stack ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Oxygen Probe Replacement Oxygen Probe Terminal Block IMPEDANCE, Ω Section TroubleshootingProbe Life Speed Fault IndicationsOfresponse Identifying and Correcting Fault Indications ThermocoupleFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 5, O2 Cell Open Fault 6, Cell Impedance Too HighFault 7, Reversed O2 Cell 11. Fault 7, Reversed O2 CellModel 3081FG Section Return of Material Model 3081FG Section Replacement PartsReplacement Parts List Index No Part Number DescriptionModel 3081FG Section Index Model 3081FG Model 3081FG Page Section Drawings and Schematics Model 3081FG Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no