Instruction Manual
Model 3081FG
f.RESET MAX CELL T
Refer to Figure
O2 CONCENTRATION
%
g.SET O2 FILTER TIME
Refer to Figure
h.TRIM 4 mA?
Refer to Figure
NOTE
Y/N (TO RESET
MAXIMUM CELL
TEMPERATURE)
26020013
Figure 3-12. RESET MAX CELL T
O2 CONCENTRATION
Before trimming the 4 mA value, you must break the loop to add the amme- ter. Power down the unit, connect the ammeter in series with Model 3081 Transmitter terminals
Press ENTER to begin editing. Use the ar- row keys to select and change the screen value to the value displayed on the installed ammeter. Press ENTER to accept the value. After the value is entered, the unit calibrates itself to ensure it outputs 4 mA. Both the display and the ammeter will dis- play 4 mA. Pressing EXIT returns to the ini- tial TRIM 4 mA? screen, and pressing NEXT displays the TRIM 20 mA? screen.
O2 CONCENTRATION
%
%
ANALYZER
ACCESS CODE
| 26020047 |
| 26020014 |
Figure | Figure |
Rosemount Analytical Inc. A Division of Emerson Process Management |