MODEL 3081 TRANSMITTER
TERMINAL BLOCK
RL 250
ANALOG OUTPUT DEVICE
LOOP CONNECTORS
SERIAL PORT & BATTERY LOOP CONNECTORS | USE INTERFACE | |
00275 0013 ONLY | ||
CHARGER MUST | ||
SERIAL PORT | ||
NOT BE USED IN | ||
| ||
HAZARDOUS AREAS |
|
| HART |
| COMMUNICATOR |
| LEAD SET |
HART COMMUNICATOR |
|
REAR PANEL | 26020040 |
Figure 4-1. Signal Line Connections, > 250 Ohms Lead Resistance
a.Method 1, For Load Resistance ≥ 250
Ohms
Refer to Figure
Explosions can result in death or seri- ous injury. Do not make connections to the HART Communicator's serial port,
Using the supplied lead set, connect the HART Communicator in parallel to the Two- Wire In Situ Oxygen Analyzer. Use any wir- ing termination points in the analog output
b.Method 2, For Load Resistance < 250 Ohms
Refer to Figure
Explosions can result in death or seri- ous injury. Do not make connections to the HART Communicator's serial port,
1.At a convenient point, break the analog output
2.Plug the load resistor into the loop connectors (located on the rear panel of the HART Communicator).