Instruction Manual
Model 3081FG
% | % |
|
26020028 | 26020029 |
Figure | Figure |
e. Fault 5, O2 Cell Open | f. Fault 6, Cell Impedance Too High |
The O2 cell connection is open. The fault displays as shown in Figure
1.Allow adequate time for the oxygen probe to reach operating temperatures. Probe temperatures below approxi- mately 500°C (932°F) may result in this fault.
2.Refer to Figure
NOTE
Check the cell output voltage at the probe terminals
3.Apply low calibration check gas (0.4% O2). Measure the cell output from the O2 cell wires at the probe terminal block. The cell output should be 100 ±20 mV. If no voltage can be meas- ured, the cell is open.
4.If the O2 cell is open, replace the oxy- gen probe per paragraph
The O2 cell impedance has exceeded 100 ohms. The fault displays as shown in Figure
1.This fault is usually indicated in con- junction with Fault 5, Cell Open. Cor- recting Fault 5 should correct Fault 6.
2.If Fault 6 appears independently, the cell has degraded beyond specification.
3.If the O2 cell has become too old, re- place the oxygen probe per paragraph