Model 3081FG

Instruction Manual

IB-106-3081 Rev. 1.5 September 2002

3.For correct viewing orientation, the dis- play may be changed 90 degrees, us- ing the following procedure:

(a)Refer to Figure 2-8. Loosen the cover lock screw until the cover lock is disengaged from the knurled surface on the threaded circuit end cap.

(b)Remove the circuit end cap.

(c)Remove the three screws retaining the display board in place.

(d)Lift and rotate the display board 90 degrees either way.

(e)Reposition the display board on the standoffs. Install and tighten all three screws.

CIRCUIT END

CAP

SCREW

DISPLAY

90

O

BOARD

 

 

 

90O

HOUSING

26020061

Figure 2-8. Display Positioning Assembly

(f) Install the circuit end cap and tighten the cover lock screw to se- cure the cover lock in place.

2-3 ELECTRICAL INSTALLATION

2

All wiring must conform to local and national

 

codes.

 

For intrinsically safe applications, re-

 

fer to drawing 1400184, page 10-2 of

 

this Instruction Bulletin.

 

Disconnect and lock out power before connecting the unit to the power supply.

Install all protective equipment covers and safety ground leads after installa- tion. Failure to install covers and ground leads could result in serious injury or death.

To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, con- nection to the main electrical power supply must be made through a circuit breaker (min 10 A) which will discon- nect all current-carrying conductors during a fault situation. This circuit breaker should also include a me- chanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.

a.General

The power supply and signal wiring should be shielded. Also, make sure the signal wiring is grounded at the Model 3081 Transmitter end only. Do not ground the signal loop at more than one point. Twisted

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-9

Image 27
Contents Model 3081FG Essential Instructions Highlights of Changes Page Model 3081FG Table of ContentsProper Calibration Check Gas Flow Rate List of IllustrationsModel 3081 Transmitter Terminal Block List of TablesModel 3081FG Definitions PrefaceModel 3081FG Component Checklist of Typical System Package Contents Section Description and SpecificationsCode Mounting Adapter Stack Side Product MatrixSystem Description System ConfigurationSystem Overview ScopeSystem Features Handling the Analyzer System Considerations Transmitter Typical System InstallationProbe SpecificationsOperating Load Region Lift OFF Maximum Power Supply VoltageOhms VDCInspect Section InstallationPRE-INSTALLATION Mechanical InstallationSide View Bottom View Front ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 3081FG Installing Model 3081 Transmitter Mounting Holes Cover LockCircuit Terminal END Block TBElectrical Installation Display Positioning AssemblyModel 3081 Transmitter 4-20 mA and Signal Connections Terminal Block ConduitThermocouple CELL-WH CELL+BK Calibration Check Gas Pneumatic InstallationReference Air Package Instrument Air Reference AirFlow SET MAX Point Knob Mounting Holes Outlet ConnectionOutlet Inlet ConnectionPower UP Section Startup and OperationGeneral Proper Calibration Check Gas Flow Rate Reestablishing Proper Calibration Check GAS Flow RateMenu Tree OperationOverview DisplayProgram DiagnosticsCalcheck MenuNavigation Infrared Remote Control IRCO2 Concentration O2 Concentration Analyzer Access Code Or SecurityAnalyzer Access Code Program MenuCell T HI O2 Concentration Indicator Fault ValueO2 Concentration Upper Cell Temperature Fault Value Upper Range VALSET O2 Filter Time Reset MAX Cell TTo Reset Maximum Cell Temperature O2 Concentration Analyzer Access CodeSET HI Bottle O2 SET LO Bottle O2Concentration Parameter Range Default Model 3081 Transmitter ParametersModel 3081 Transmitter Parameters SET Code Diagnostics Menu Show FaultC mV 24. Previous Slope Previous Constant 23. Cell ImpedanceSW Build Number SW Build DateCalcheck Menu Accept LOW O2 27. in MANUAL?29. Accept LOW O2 Slope Model 3081FG Connections Hart Communicator Signal LineSection HART/AMS OverviewMethod 2, For Load Resistance 250 Ohms Analog Output Device Loop ConnectorsMenu Tree for Hart Communicator TWO-WIRE in Situ Oxygen AnalyzerHart Communicator PC OFF-LINE and ON-LINE OperationsDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 3081FG PC Board Stack Replacement Section Maintenance and ServiceModel 3081 Electronics Replacement Display Board ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Oxygen Probe Replacement Oxygen Probe Terminal Block Section Troubleshooting Probe LifeIMPEDANCE, Ω Fault Indications OfresponseSpeed Thermocouple Identifying and Correcting Fault IndicationsFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 6, Cell Impedance Too High Fault 5, O2 Cell Open11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 CellModel 3081FG Section Return of Material Model 3081FG Index No Part Number Description Replacement PartsReplacement Parts List SectionModel 3081FG Section Index Model 3081FG Model 3081FG Page Section Drawings and Schematics Model 3081FG Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no