Model 3081FG
Instruction Manual
September 2002
1
1-3 SPECIFICATIONS
Net O2 Range | 0 to 25% O2 |
Fully Field Selectable via the HART Interface |
|
Lowest Limit | 0.05% O2 |
Highest Limit | 25.00% O2 |
Accuracy | ±1.5% of reading or 0.05% O2, whichever is greater |
System Response to Calibration Check Gas | Initial response in less than 3 seconds |
PROBE | T90 in less than 10 seconds |
| |
Lengths | 20 in. (508 mm) |
| 26 in. (660 mm) |
Temperature Limits | 34.625 in. (880 mm) |
| |
Process Temperature Limits | 550° to 1400°C (1022° to 2552°F) |
| Operation to 1600°C (2912°F) with reduced cell life. |
Ambient | |
Mounting and Mounting Position | Vertical or Horizontal |
Materials of Construction |
|
Process Wetted Parts |
|
Inner Probe | Zirconia |
Outer Protection Tube | Alumina [1600°C (2912°F) limit] |
| Inconel 600 [1000°C (1832°F) limit] |
Probe Junction Box | Cast aluminum |
Speed of Installation/Withdrawal | 1 in. (25.4 mm) per minute |
Hazardous Area Certification | Intrinsically safe per EN50 014 (1977), clause 1.3(1) |
Reference Air Requirement | 100 ml per minute (0.2 scfh) of clean, dry instrument |
| air; 1/4 in. tube fittings |
Calibration Check Gas Fittings | 1/4 in. tube fittings |
Cabling | Two twisted pairs, shielded |
(1)Thermocouple and O2 probe cell are both unpowered, developing a millivolt emf, and are considered a “simple apparatus” by certifying agencies.
Rosemount Analytical Inc. A Division of Emerson Process Management | Description and Specifications |