Emerson 3081FG instruction manual Cell Impedance, Previous Slope Previous Constant

Page 42

Instruction Manual

IB-106-3081 Rev. 1.5 September 2002

Model 3081FG

O2 CONCENTRATION

O2 CONCENTRATION

%

%

O2 CELL

IMPEDANCE STATUS

26020020

Figure 3-23. CELL IMPEDANCE

d.CELL IMPEDANCE

Refer to Figure 3-23. Use this screen to ex- amine the O2 cell impedance status. GOOD indicates the cell is operating normally.

WARN indicates the cell has degraded but is still operational. HI indicates that the cell has degraded but is still operational; how- ever, failure will occur soon. Pressing NEXT accesses the PREVIOUS SLOPE screen.

CALCHECK CELL

SLOPE VALUE

26020052

Figure 3-24. PREVIOUS SLOPE

f.PREVIOUS CONSTANT

Refer to Figure 3-25. Use this screen to ex- amine the cell zero constant calculated from the most recent calibration check. The con- stant represents the voltage generated by the cell when no difference exists between the amount of O2 on the reference and pro- cess sides of the cell. Press NEXT to ac- cess the MAX CELL T screen.

NOTE

Temperature influences cell imped- ance. Wait until the cell is at operating temperature before checking cell im- pedance. If checked before the cell reaches operating temperature [550°C (1022°F)], this screen displays a fail indication.

e.PREVIOUS SLOPE

Refer to Figure 3-24. Use this screen to ex- amine the slope calculated from the most recent calibration check. The slope is the amount of cell voltage generated for a given O2 value. For each calibration check, record the slope over the life of the probe. Tracking the slope will indicate if the probe is de- grading. Press NEXT to access the PREVI- OUS CONSTANT screen.

O2 CONCENTRATION

%

CALCHECK CELL

CONSTANT VALUE

26020053

Figure 3-25. PREVIOUS CONSTANT

3-12 Startup and Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 42
Contents Model 3081FG Essential Instructions Highlights of Changes Page Table of Contents Model 3081FGList of Illustrations Proper Calibration Check Gas Flow RateList of Tables Model 3081 Transmitter Terminal BlockModel 3081FG Preface DefinitionsModel 3081FG Section Description and Specifications Component Checklist of Typical System Package ContentsProduct Matrix Code Mounting Adapter Stack SideScope System ConfigurationSystem Overview System DescriptionSystem Features Handling the Analyzer System Considerations Typical System Installation TransmitterSpecifications ProbeVDC Lift OFF Maximum Power Supply VoltageOhms Operating Load RegionMechanical Installation Section InstallationPRE-INSTALLATION InspectFront View Side View Bottom ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 3081FG Installing Model 3081 Transmitter Block TB Cover LockCircuit Terminal END Mounting HolesDisplay Positioning Assembly Electrical InstallationModel 3081 Transmitter 4-20 mA and Signal Connections Terminal Block ConduitThermocouple CELL-WH CELL+BK Instrument Air Reference Air Pneumatic InstallationReference Air Package Calibration Check GasInlet Connection Outlet ConnectionOutlet Flow SET MAX Point Knob Mounting HolesPower UP Section Startup and OperationGeneral Reestablishing Proper Calibration Check GAS Flow Rate Proper Calibration Check Gas Flow RateDisplay OperationOverview Menu TreeMenu DiagnosticsCalcheck ProgramInfrared Remote Control IRC NavigationProgram Menu O2 Concentration Analyzer Access Code Or SecurityAnalyzer Access Code O2 ConcentrationUpper Range VAL O2 Concentration Indicator Fault ValueO2 Concentration Upper Cell Temperature Fault Value Cell T HIO2 Concentration Analyzer Access Code Reset MAX Cell TTo Reset Maximum Cell Temperature SET O2 Filter TimeSET HI Bottle O2 SET LO Bottle O2Concentration Parameter Range Default Model 3081 Transmitter ParametersModel 3081 Transmitter Parameters SET Code Diagnostics Menu Show FaultC mV 23. Cell Impedance 24. Previous Slope Previous ConstantSW Build Number SW Build DateCalcheck Menu 27. in MANUAL? Accept LOW O229. Accept LOW O2 Slope Model 3081FG Overview Hart Communicator Signal LineSection HART/AMS ConnectionsAnalog Output Device Loop Connectors Method 2, For Load Resistance 250 OhmsOFF-LINE and ON-LINE Operations TWO-WIRE in Situ Oxygen AnalyzerHart Communicator PC Menu Tree for Hart CommunicatorDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 3081FG Display Board Replacement Section Maintenance and ServiceModel 3081 Electronics Replacement PC Board Stack ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Oxygen Probe Replacement Oxygen Probe Terminal Block Section Troubleshooting Probe LifeIMPEDANCE, Ω Fault Indications OfresponseSpeed Identifying and Correcting Fault Indications ThermocoupleFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 5, O2 Cell Open Fault 6, Cell Impedance Too HighFault 7, Reversed O2 Cell 11. Fault 7, Reversed O2 CellModel 3081FG Section Return of Material Model 3081FG Section Replacement PartsReplacement Parts List Index No Part Number DescriptionModel 3081FG Section Index Model 3081FG Model 3081FG Page Section Drawings and Schematics Model 3081FG Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no