Instruction Manual

Model 3081FG

IB-106-3081 Rev. 1.5 September 2002

SECTION 2

INSTALLATION

2-1 PRE-INSTALLATION

a.Inspect

Carefully inspect the shipping container for any evidence of damage. If the container is damaged, notify the carrier immediately.

b.Packing List

Confirm that all items shown on the packing list are present. Notify Rosemount Analytical immediately if items are missing.

Before installing this equipment, read the “Safety instructions for the wiring and installation of this apparatus” at the front of this Instruction Bulletin. Failure to follow the safety instruc- tions could result in serious injury or death.

2-2 MECHANICAL INSTALLATION

Avoid installation locations near steam soot blowers.

a.Locating Oxygen Probe

1.The location of the oxygen probe in the stack or flue is important for maxi- mum accuracy in the oxygen analyzing process. The probe must be positioned so the gas it measures is representa- tive of the process. Best results are normally obtained if the probe is posi- tioned near the center of the duct (40- 60% insertion). Longer ducts may re- quire several analyzers since the O2 can vary due to stratification. A point too near the wall of the duct, or the in- side radius of a bend, may not provide a representative sample because of the very low flow conditions. The

2

sensing point should be selected so the process gas temperature falls within a range of 550° to 1600°C (1022° to 2912°F). Figure 2-1provides mechanical installation references.

2.Check the flue or stack for holes and air leakage. The presence of this con- dition will substantially affect the accu- racy of the oxygen reading. Therefore, either make the necessary repairs or install the probe upstream of any leakage.

3.Ensure the area is clear of internal and external obstructions that will interfere with installation and maintenance ac- cess to the probe. Allow adequate clearance for probe removal (Figure 2-1).

b.Installing Oxygen Probe

The probe was specially packaged to prevent breakage due to handling. Do not remove the padding material from the probe until immediately before installation.

1.Ensure all components are available to install the probe.

NOTE

Leave the probe inner protective cover in place until installation. This is re- quired to protect the ceramic cell dur- ing movement.

2.If using an optional adapter plate (Figure 2-2)or an optional mounting flange (Figure 2-3),weld or bolt the component onto the duct. The through hole in the stack or duct wall and re- fractory material must be 2 in. (50.8 mm) diameter, minimum.

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-1

Image 19
Contents Model 3081FG Essential Instructions Highlights of Changes Page Model 3081FG Table of ContentsProper Calibration Check Gas Flow Rate List of IllustrationsModel 3081 Transmitter Terminal Block List of TablesModel 3081FG Definitions PrefaceModel 3081FG Component Checklist of Typical System Package Contents Section Description and SpecificationsCode Mounting Adapter Stack Side Product MatrixSystem Description System ConfigurationSystem Overview ScopeSystem Features Handling the Analyzer System Considerations Transmitter Typical System InstallationProbe SpecificationsOperating Load Region Lift OFF Maximum Power Supply VoltageOhms VDCInspect Section InstallationPRE-INSTALLATION Mechanical InstallationSide View Bottom View Front ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 3081FG Installing Model 3081 Transmitter Mounting Holes Cover LockCircuit Terminal END Block TBElectrical Installation Display Positioning AssemblyTerminal Block Conduit Model 3081 Transmitter 4-20 mA and Signal ConnectionsThermocouple CELL-WH CELL+BK Calibration Check Gas Pneumatic InstallationReference Air Package Instrument Air Reference AirFlow SET MAX Point Knob Mounting Holes Outlet ConnectionOutlet Inlet ConnectionSection Startup and Operation Power UPGeneral Proper Calibration Check Gas Flow Rate Reestablishing Proper Calibration Check GAS Flow RateMenu Tree OperationOverview DisplayProgram DiagnosticsCalcheck MenuNavigation Infrared Remote Control IRCO2 Concentration O2 Concentration Analyzer Access Code Or SecurityAnalyzer Access Code Program MenuCell T HI O2 Concentration Indicator Fault ValueO2 Concentration Upper Cell Temperature Fault Value Upper Range VALSET O2 Filter Time Reset MAX Cell TTo Reset Maximum Cell Temperature O2 Concentration Analyzer Access CodeSET LO Bottle O2 SET HI Bottle O2Concentration Model 3081 Transmitter Parameters Parameter Range DefaultModel 3081 Transmitter Parameters SET Code Show Fault Diagnostics MenuC mV 24. Previous Slope Previous Constant 23. Cell ImpedanceSW Build Date SW Build NumberCalcheck Menu Accept LOW O2 27. in MANUAL?29. Accept LOW O2 Slope Model 3081FG Connections Hart Communicator Signal LineSection HART/AMS OverviewMethod 2, For Load Resistance 250 Ohms Analog Output Device Loop ConnectorsMenu Tree for Hart Communicator TWO-WIRE in Situ Oxygen AnalyzerHart Communicator PC OFF-LINE and ON-LINE OperationsDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 3081FG PC Board Stack Replacement Section Maintenance and ServiceModel 3081 Electronics Replacement Display Board ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Oxygen Probe Replacement Oxygen Probe Terminal Block Probe Life Section TroubleshootingIMPEDANCE, Ω Ofresponse Fault IndicationsSpeed Thermocouple Identifying and Correcting Fault IndicationsFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 6, Cell Impedance Too High Fault 5, O2 Cell Open11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 CellModel 3081FG Section Return of Material Model 3081FG Index No Part Number Description Replacement PartsReplacement Parts List SectionModel 3081FG Section Index Model 3081FG Model 3081FG Page Section Drawings and Schematics Model 3081FG Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no