Emerson 3081FG instruction manual Metal Wall Masonry Wall Stack or Duct

Page 21

 

 

 

Instruction Manual

 

Model 3081FG

 

 

IB-106-3081 Rev. 1.5

 

 

 

September 2002

 

 

 

METAL WALL

MASONRY WALL

 

 

 

STACK OR DUCT

STACK

 

A

 

 

WELD PIPE TO

2

 

 

 

ADAPTER PLATE

B

 

 

 

 

 

A

2.50 (63.5)

3.50 (89)

 

 

MIN. DIA.

O.D. REF

 

 

 

 

C

 

 

 

 

NOTE: DIMENSIONS ARE

 

 

JOINT MUST

 

IN INCHES WITH

 

BE AIR TIGHT

 

MILLIMETERS IN

 

WELD OR BOLT ADAPTER

3.00 SCHEDULE 40

 

PARENTHESES.

PLATE TO STACK OR DUCT.

PIPE SLEEVE

 

 

JOINT MUST BE AIR TIGHT.

SUPPLIED BY CUSTOMER

 

 

 

 

PLATE DIMENSIONS

 

ANSI

DIN

JIS

DIMENSION

4512C34G01

4512C36G01

4512C35G01

 

 

 

 

 

 

 

“A”

6.00 (153)

7.5 (191)

6.50 (165)

 

 

 

 

“B” THREAD

0.625-11

M-16x2

M-12x1.75

 

 

 

 

“C” DIA.

4.75 (121)

5.71 (145)

5.12 (130)

29750002

Figure 2-2. Optional Adapter Plate

TAP 1.25 NPT

0.50 (12.7)

B

A

C

 

 

 

 

 

 

 

NOTE: DIMENSIONS ARE IN

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INCHES WITH MILLIMETERS

 

 

 

 

 

 

 

 

IN PARENTHESES.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FLANGE DIMENSIONS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ANSI

DIN

 

JIS

MODEL 450

DIMENSION

5R10158H01

5R10158H02

5R10158H03

5R10158H04

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

“A”

DIA.

6.00 (153)

7.28 (185)

 

6.10 (155)

9.00 (229)

 

 

 

 

 

 

 

 

 

“B”

DIA.

0.75 (20)

0.71 (18)

 

0.59 (15)

0.50 (13)

 

 

 

 

 

 

 

 

 

“C”

DIA.

4.75 (121)

5.71 (145)

 

5.12 (130)

7.68 (195)

 

29750003

Figure 2-3. Optional Probe Mounting Flange

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-3

Image 21
Contents Model 3081FG Essential Instructions Highlights of Changes Page Model 3081FG Table of ContentsProper Calibration Check Gas Flow Rate List of IllustrationsModel 3081 Transmitter Terminal Block List of TablesModel 3081FG Definitions PrefaceModel 3081FG Component Checklist of Typical System Package Contents Section Description and SpecificationsCode Mounting Adapter Stack Side Product MatrixSystem Overview System ConfigurationScope System DescriptionSystem Features Handling the Analyzer System Considerations Transmitter Typical System InstallationProbe SpecificationsOhms Lift OFF Maximum Power Supply VoltageVDC Operating Load RegionPRE-INSTALLATION Section InstallationMechanical Installation InspectSide View Bottom View Front ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 3081FG Installing Model 3081 Transmitter Circuit Terminal END Cover LockBlock TB Mounting HolesElectrical Installation Display Positioning AssemblyModel 3081 Transmitter 4-20 mA and Signal Connections Terminal Block ConduitThermocouple CELL-WH CELL+BK Reference Air Package Pneumatic InstallationInstrument Air Reference Air Calibration Check GasOutlet Outlet ConnectionInlet Connection Flow SET MAX Point Knob Mounting HolesPower UP Section Startup and OperationGeneral Proper Calibration Check Gas Flow Rate Reestablishing Proper Calibration Check GAS Flow RateOverview OperationDisplay Menu TreeCalcheck DiagnosticsMenu ProgramNavigation Infrared Remote Control IRCAnalyzer Access Code O2 Concentration Analyzer Access Code Or SecurityProgram Menu O2 ConcentrationO2 Concentration Upper Cell Temperature Fault Value O2 Concentration Indicator Fault ValueUpper Range VAL Cell T HITo Reset Maximum Cell Temperature Reset MAX Cell TO2 Concentration Analyzer Access Code SET O2 Filter TimeSET HI Bottle O2 SET LO Bottle O2Concentration Parameter Range Default Model 3081 Transmitter ParametersModel 3081 Transmitter Parameters SET Code Diagnostics Menu Show FaultC mV 24. Previous Slope Previous Constant 23. Cell ImpedanceSW Build Number SW Build DateCalcheck Menu Accept LOW O2 27. in MANUAL?29. Accept LOW O2 Slope Model 3081FG Section HART/AMS Hart Communicator Signal LineOverview ConnectionsMethod 2, For Load Resistance 250 Ohms Analog Output Device Loop ConnectorsHart Communicator PC TWO-WIRE in Situ Oxygen AnalyzerOFF-LINE and ON-LINE Operations Menu Tree for Hart CommunicatorDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 3081FG Model 3081 Electronics Replacement Section Maintenance and ServiceDisplay Board Replacement PC Board Stack ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Oxygen Probe Replacement Oxygen Probe Terminal Block Section Troubleshooting Probe LifeIMPEDANCE, Ω Fault Indications OfresponseSpeed Thermocouple Identifying and Correcting Fault IndicationsFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 6, Cell Impedance Too High Fault 5, O2 Cell Open11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 CellModel 3081FG Section Return of Material Model 3081FG Replacement Parts List Replacement PartsSection Index No Part Number DescriptionModel 3081FG Section Index Model 3081FG Model 3081FG Page Section Drawings and Schematics Model 3081FG Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no