Emerson 3081FG instruction manual Horizontal Probe Installation

Page 22

Instruction Manual

IB-106-3081 Rev. 1.5 September 2002

Model 3081FG

STACK OR DUCT

METAL WALL

WELD PIPE TO

METAL WALL

2.0IN. (51 mm) MIN. DIA.

REFRACTORY

2 IN. NPT

1.25 NPT

CUSTOMER

SUPPLIED

ADAPTER

SCHEDULE 40

PIPE

 

 

 

STACK OR DUCT

 

 

 

 

 

METAL WALL

INSULATE IF EXPOSED

 

 

 

 

 

 

ADAPTER

TO AMBIENT WEATHER

 

 

 

 

CONDITIONS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2 IN. NPT

SCHEDULE 40

PIPE

CALIBRATION

SYSTEM

CHECK

GAS LINE

CABLE

REFERENCE

 

AIR LINE

29750004

Figure 2-4. Horizontal Probe Installation

3.If the optional adapter plates are not used, a 2 in. NPT, schedule 40, pipe nipple (Figure 2-4) should be welded to the stack or duct wall.

When a 2 in. NPT to 1.25 NPT adapter is threaded to the welded pipe nipple,

the adapter provides the pipe threads needed for the probe’s process fitting.

4.Where high particulate or slag is in the flue gas stream, it may be desirable to inset the probe in the refractory as shown in Figure 2-5.Use pipe cou- plings and nipples to adjust the probe insertion depth.

2-4 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 22
Contents Model 3081FG Essential Instructions Highlights of Changes Page Table of Contents Model 3081FGList of Illustrations Proper Calibration Check Gas Flow RateList of Tables Model 3081 Transmitter Terminal BlockModel 3081FG Preface DefinitionsModel 3081FG Section Description and Specifications Component Checklist of Typical System Package ContentsProduct Matrix Code Mounting Adapter Stack SideScope System ConfigurationSystem Overview System DescriptionSystem Features Handling the Analyzer System Considerations Typical System Installation TransmitterSpecifications ProbeVDC Lift OFF Maximum Power Supply VoltageOhms Operating Load RegionMechanical Installation Section InstallationPRE-INSTALLATION InspectFront View Side View Bottom ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 3081FG Installing Model 3081 Transmitter Block TB Cover LockCircuit Terminal END Mounting HolesDisplay Positioning Assembly Electrical InstallationTerminal Block Conduit Model 3081 Transmitter 4-20 mA and Signal ConnectionsThermocouple CELL-WH CELL+BK Instrument Air Reference Air Pneumatic InstallationReference Air Package Calibration Check GasInlet Connection Outlet ConnectionOutlet Flow SET MAX Point Knob Mounting HolesSection Startup and Operation Power UPGeneral Reestablishing Proper Calibration Check GAS Flow Rate Proper Calibration Check Gas Flow RateDisplay OperationOverview Menu TreeMenu DiagnosticsCalcheck ProgramInfrared Remote Control IRC NavigationProgram Menu O2 Concentration Analyzer Access Code Or SecurityAnalyzer Access Code O2 ConcentrationUpper Range VAL O2 Concentration Indicator Fault ValueO2 Concentration Upper Cell Temperature Fault Value Cell T HIO2 Concentration Analyzer Access Code Reset MAX Cell TTo Reset Maximum Cell Temperature SET O2 Filter TimeSET LO Bottle O2 SET HI Bottle O2Concentration Model 3081 Transmitter Parameters Parameter Range DefaultModel 3081 Transmitter Parameters SET Code Show Fault Diagnostics MenuC mV 23. Cell Impedance 24. Previous Slope Previous ConstantSW Build Date SW Build NumberCalcheck Menu 27. in MANUAL? Accept LOW O229. Accept LOW O2 Slope Model 3081FG Overview Hart Communicator Signal LineSection HART/AMS ConnectionsAnalog Output Device Loop Connectors Method 2, For Load Resistance 250 OhmsOFF-LINE and ON-LINE Operations TWO-WIRE in Situ Oxygen AnalyzerHart Communicator PC Menu Tree for Hart CommunicatorDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 3081FG Display Board Replacement Section Maintenance and ServiceModel 3081 Electronics Replacement PC Board Stack ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Oxygen Probe Replacement Oxygen Probe Terminal Block Probe Life Section TroubleshootingIMPEDANCE, Ω Ofresponse Fault IndicationsSpeed Identifying and Correcting Fault Indications ThermocoupleFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 5, O2 Cell Open Fault 6, Cell Impedance Too HighFault 7, Reversed O2 Cell 11. Fault 7, Reversed O2 CellModel 3081FG Section Return of Material Model 3081FG Section Replacement PartsReplacement Parts List Index No Part Number DescriptionModel 3081FG Section Index Model 3081FG Model 3081FG Page Section Drawings and Schematics Model 3081FG Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no