Emerson 3081FG instruction manual Accept LOW O2 Slope

Page 45

Model 3081FG

Instruction Manual

IB-106-3081 Rev. 1.5 September 2002

O2 CONCENTRATION

%

%

LOW CALIBRATION

CHECK GAS O2 VALUE

26020056

Figure 3-29. ACCEPT LOW O2

d.SLOPE

Refer to Figure 3-30. Use this screen to ex- amine the slope calculated from current calibration check. The slope is the amount of cell voltage generated for a given O2 value. After each calibration check, record the slope over the life of the probe. Tracking the slope will indicate if the probe is de- grading. Press NEXT to access the CON- STANT screen.

O2 CONCENTRATION

%

CALCHECK CELL

SLOPE VALUE

26020057

Figure 3-30. SLOPE

e.CONSTANT

Refer to Figure 3-31. Use this screen to ex- amine the cell zero constant calculated from the current calibration check. The constant represents the voltage generated by the cell when no difference exists between the amount of O2 on the reference and process

sides of the cell. Note this value for com-

parison against future calibration checks.3 Press RESET or EXIT to return to the

PROCESS DISPLAY screen.

O2 CONCENTRATION

%

CALCHECK CELL

CONSTANT VALUE

26020058

Figure 3-31. CONSTANT

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation 3-15

Image 45
Contents Model 3081FG Essential Instructions Highlights of Changes Page Model 3081FG Table of ContentsProper Calibration Check Gas Flow Rate List of IllustrationsModel 3081 Transmitter Terminal Block List of TablesModel 3081FG Definitions PrefaceModel 3081FG Component Checklist of Typical System Package Contents Section Description and SpecificationsCode Mounting Adapter Stack Side Product MatrixSystem Overview System ConfigurationScope System DescriptionSystem Features Handling the Analyzer System Considerations Transmitter Typical System InstallationProbe SpecificationsOhms Lift OFF Maximum Power Supply VoltageVDC Operating Load RegionPRE-INSTALLATION Section InstallationMechanical Installation InspectSide View Bottom View Front ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 3081FG Installing Model 3081 Transmitter Circuit Terminal END Cover LockBlock TB Mounting HolesElectrical Installation Display Positioning AssemblyModel 3081 Transmitter 4-20 mA and Signal Connections Terminal Block ConduitThermocouple CELL-WH CELL+BK Reference Air Package Pneumatic InstallationInstrument Air Reference Air Calibration Check GasOutlet Outlet ConnectionInlet Connection Flow SET MAX Point Knob Mounting HolesPower UP Section Startup and OperationGeneral Proper Calibration Check Gas Flow Rate Reestablishing Proper Calibration Check GAS Flow RateOverview OperationDisplay Menu TreeCalcheck DiagnosticsMenu ProgramNavigation Infrared Remote Control IRCAnalyzer Access Code O2 Concentration Analyzer Access Code Or SecurityProgram Menu O2 ConcentrationO2 Concentration Upper Cell Temperature Fault Value O2 Concentration Indicator Fault ValueUpper Range VAL Cell T HITo Reset Maximum Cell Temperature Reset MAX Cell TO2 Concentration Analyzer Access Code SET O2 Filter TimeSET HI Bottle O2 SET LO Bottle O2Concentration Parameter Range Default Model 3081 Transmitter ParametersModel 3081 Transmitter Parameters SET Code Diagnostics Menu Show FaultC mV 24. Previous Slope Previous Constant 23. Cell ImpedanceSW Build Number SW Build DateCalcheck Menu Accept LOW O2 27. in MANUAL?29. Accept LOW O2 Slope Model 3081FG Section HART/AMS Hart Communicator Signal LineOverview ConnectionsMethod 2, For Load Resistance 250 Ohms Analog Output Device Loop ConnectorsHart Communicator PC TWO-WIRE in Situ Oxygen AnalyzerOFF-LINE and ON-LINE Operations Menu Tree for Hart CommunicatorDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 3081FG Model 3081 Electronics Replacement Section Maintenance and ServiceDisplay Board Replacement PC Board Stack ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Oxygen Probe Replacement Oxygen Probe Terminal Block Section Troubleshooting Probe LifeIMPEDANCE, Ω Fault Indications OfresponseSpeed Thermocouple Identifying and Correcting Fault IndicationsFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 6, Cell Impedance Too High Fault 5, O2 Cell Open11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 CellModel 3081FG Section Return of Material Model 3081FG Replacement Parts List Replacement PartsSection Index No Part Number DescriptionModel 3081FG Section Index Model 3081FG Model 3081FG Page Section Drawings and Schematics Model 3081FG Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no