ResMed Mirage Activa, Ultra Mirage manual Option 3 Sterrad Sterilization System, Pre-sterilization

Page 5

Clinical Use Only

Option 3: STERRAD Sterilization System

Note: The Ultra Mirage™ II Nasal Mask is not validated for the STERRAD Sterilization System.

Pre-sterilization

1Disassemble the mask.

2Clean the individual parts of the mask with a soft bristle brush for one minute while soaking them in detergent (eg, Alconox). Pay particular attention to all crevices and cavities.

3Rinse the components twice by agitating them vigorously in drinking quality water and allow them to air dry out of direct sunlight.

ResMed has tested the detergent Alconox (diluted at 1%).

Sterilization

Sterilize the sterilizable mask components as described in the manufacturer’s instructions for the STERRAD Sterilization System (eg, Sterrad 100S or Sterrad NX).

Testing by ResMed has shown that the mask components can withstand 15 cycles of the STERRAD 100S Sterilization System, when using the tray method.

Note: The use of pouches is not recommended.

Inspection after Disinfection/Sterilization

Perform a visual inspection of each mask component. If any visible deterioration of a mask component is apparent (cracking, crazing, tears etc), the mask component should be discarded and replaced. Slight discoloration of the silicone components may occur and is acceptable.

Note: If a healthcare facility requires an additional disinfection or sterilization process after reassembly, the number of validated cycles is halved.

Parts not Validated for Multipatient Use

Some parts of the masks have not been validated for multipatient use. These parts need to be replaced with new parts when using between patients. Please check the Components Card on the website www.resmed.com for a list of available replacement parts for each mask system.

4

Image 5
Contents ResMed Masks English Disinfection Option 1 Chemical DisinfectionPre-disinfection Regional Requirements Option 2 Thermal DisinfectionEuropean Standards Inspection after Disinfection/Sterilization Option 3 Sterrad Sterilization SystemParts not Validated for Multipatient Use Pre-sterilizationClinical Use Only Clinical Use Only Deutsch Warnungen UND VorsichtsmassnahmenDesinfektion Option 1 Chemische DesinfektionVordesinfektion Regionale Anforderungen Option 2 Thermale DesinfektionEuropäische Normen Inspektion nach der Desinfektion/Sterilisation Option 3 Sterrad SterilisationssystemVorsterilisation SterilisationMirage Full Face Maske NV ohne Luftauslassöffnungen Serie Nur zur klinischen Behandlung Français Avertissements ET PrecautionsDésinfection Option 1 Désinfection chimiquePré-désinfection Exigences régionales Option 2 Désinfection thermiqueNormes européennes Inspection après la désinfection/stérilisation Option 3 Système de stérilisation SterradPré-stérilisation StérilisationEnsemble tuyau à air court Utilisation clinique uniquement Italiano Avvertenze E PrecauzioniDisinfezione Opzione 1 Disinfezione chimicaPre-disinfezione Requisiti regionali Opzione 2 Disinfezione termicaStandard europei Ispezione dopo la disinfezione/sterilizzazione Opzione 3 Sistema di sterilizzazione SterradParti non approvate per l’uso multipaziente Pre-sterilizzazioneNessuna Solo per uso clinico Español Advertencias Y PrecaucionesDesinfección Opción 1 desinfección químicaPrevio a la desinfección Requisitos regionales Opción 2 desinfección térmicaNormas europeas Inspección posterior a la desinfección o esterilización Opción 3 Sistema de esterilización SterradPiezas no validadas para el uso entre pacientes Previo a la esterilizaciónSistema de Almohadillas Nasales Mirage Swift Solamente para uso clínico Português Avisos E PrecauçõesDesinfecção Opção 1 Desinfecção QuímicaLimpeza antes da desinfecção Requisitos Regionais Opção 2 Desinfecção TérmicaNormas Europeias Inspecção após desinfecção/esterilização Opção 3 Sistema de Esterilização SterradLimpeza antes da esterilização EsterilizaçãoApenas para uso clínico Apenas para uso clínico Nederlands Waarschuwingen EN OpmerkingenDesinfectie Optie 1 Chemische desinfectieVoorafgaand aan de desinfectie Voorafgaand aan desinfectie Optie 2 Thermische desinfectieEuropese Normen Regionale vereistenInspectie na desinfectie/sterilisatie Optie 3 Sterrad SterilisatiesysteemVoorafgaand aan sterilisatie SterilisatieAlleen voor gebruik in de ziekenhuissituatie Alleen voor gebruik in de ziekenhuissituatie