ResMed Mirage Swift, Ultra Mirage II, Mirage Vista, Mirage Kidsta manual Clinical Use Only

Page 6

Clinical Use Only

The parts that cannot be disinfected or sterilized are listed in the following table.

Masks/Pillows System

Parts that cannot be disinfected or

 

 

sterilized

 

 

 

•

Mirage Vista™ Nasal Mask

Swivel, short tubing

•

Mirage Activa™ Nasal Mask

 

• Mirage Kidsta™ Nasal Mask

 

 

 

 

•

Mirage Swift™ Nasal Pillows System

Short tube assembly

• Mirage Swift™ II Nasal Pillows System

 

 

 

• Mirage™ Full Face Mask Series 2

Valve membrane

• Ultra Mirage™ Full Face Mask

 

 

 

 

•

Non-vented Ultra Mirage™ Nasal Mask

None

• Non-vented Ultra Mirage™ Full Face Mask

 

• Ultra Mirage™ II Nasal Mask

 

• Non-vented Mirage™ Full Face Mask Series 2

 

 

 

 

Note: Parts that cannot be disinfected or sterilized should be replaced when using between patients.

Note: Different disinfectants and detergents are available in each country. For this reason, ResMed has identified alternatives as examples and has validated the disinfectants and detergents listed in this guide.

English

5

Image 6
Contents ResMed Masks English Option 1 Chemical Disinfection Pre-disinfectionDisinfection Option 2 Thermal Disinfection European StandardsRegional Requirements Parts not Validated for Multipatient Use Option 3 Sterrad Sterilization SystemInspection after Disinfection/Sterilization Pre-sterilizationClinical Use Only Clinical Use Only Warnungen UND Vorsichtsmassnahmen DeutschOption 1 Chemische Desinfektion VordesinfektionDesinfektion Option 2 Thermale Desinfektion Europäische NormenRegionale Anforderungen Vorsterilisation Option 3 Sterrad SterilisationssystemInspektion nach der Desinfektion/Sterilisation SterilisationMirage Full Face Maske NV ohne Luftauslassöffnungen Serie Nur zur klinischen Behandlung Avertissements ET Precautions FrançaisOption 1 Désinfection chimique Pré-désinfectionDésinfection Option 2 Désinfection thermique Normes européennesExigences régionales Pré-stérilisation Option 3 Système de stérilisation SterradInspection après la désinfection/stérilisation StérilisationEnsemble tuyau à air court Utilisation clinique uniquement Avvertenze E Precauzioni ItalianoOpzione 1 Disinfezione chimica Pre-disinfezioneDisinfezione Opzione 2 Disinfezione termica Standard europeiRequisiti regionali Parti non approvate per l’uso multipaziente Opzione 3 Sistema di sterilizzazione SterradIspezione dopo la disinfezione/sterilizzazione Pre-sterilizzazioneNessuna Solo per uso clinico Advertencias Y Precauciones EspañolOpción 1 desinfección química Previo a la desinfecciónDesinfección Opción 2 desinfección térmica Normas europeasRequisitos regionales Piezas no validadas para el uso entre pacientes Opción 3 Sistema de esterilización SterradInspección posterior a la desinfección o esterilización Previo a la esterilizaciónSistema de Almohadillas Nasales Mirage Swift Solamente para uso clínico Avisos E Precauções PortuguêsOpção 1 Desinfecção Química Limpeza antes da desinfecçãoDesinfecção Opção 2 Desinfecção Térmica Normas EuropeiasRequisitos Regionais Limpeza antes da esterilização Opção 3 Sistema de Esterilização SterradInspecção após desinfecção/esterilização EsterilizaçãoApenas para uso clínico Apenas para uso clínico Waarschuwingen EN Opmerkingen NederlandsOptie 1 Chemische desinfectie Voorafgaand aan de desinfectieDesinfectie Europese Normen Optie 2 Thermische desinfectieVoorafgaand aan desinfectie Regionale vereistenVoorafgaand aan sterilisatie Optie 3 Sterrad SterilisatiesysteemInspectie na desinfectie/sterilisatie SterilisatieAlleen voor gebruik in de ziekenhuissituatie Alleen voor gebruik in de ziekenhuissituatie