MINOR TROUBLESHOOTING

The following troubleshooting chart will help you analyze and correct minor oxygen concentrator malfunctions. If the suggested procedures do not help, switch to your reserve oxygen sys- tem and call your Sunrise Medical homecare provider. Do not attempt any other maintenance.

WARNING–To avoid electric shock hazard, do not remove the cabinet.The cabinet should only be removed by a qualified Sunrise Medical homecare technician.

MINOR TROUBLESHOOTING CHART

 

Symptom

 

Possible Cause

 

Remedy

 

 

 

 

 

 

 

A. Unit does not operate. Power light is off when the power

1.

Power cord not properly inserted into wall outlet.

1.

Check power cord connection at the wall outlet. On 230 volt units,

 

 

switch is “On.” Audible alarm is pulsing and Service Required

 

 

 

also check the back of the unit.

 

 

light is flashing.

2.

No power at wall outlet.

2.

Check your home circuit breaker and reset if necessary. Use a

 

 

 

 

 

 

different wall outlet if the situation occurs again.

 

 

 

3.

Oxygen concentrator circuit breaker activated

3.

Press the concentrator circuit breaker reset button located below

 

 

 

 

(only 115 volt units).

 

the power switch (only 115 volt units). Use a different wall outlet if

 

 

 

 

 

 

the situation occurs again.

 

 

 

 

 

 

If the above remedies do not work, contact your

 

 

 

 

 

 

Sunrise Medical provider.

 

 

 

 

 

 

 

 

 

 

 

B. (DeVilbiss 3 & 5 Liter Series) Unit does not operate but the Power

1.

Air filter is blocked.

1.

Check the air filter. If the filter is dirty, wash it following the

 

 

light is on when the power switch is “On.” Red Service

 

 

 

cleaning instructions on page 18.

 

 

Required light is illuminated.Audible alarm may be sounding.

2.

Exhaust is blocked.

2.

Check the exhaust area; make sure there is nothing restricting the

 

 

 

 

 

 

unit exhaust.

 

 

 

 

 

 

If the above remedies do not work, contact your

 

 

 

 

 

 

Sunrise Medical provider.

 

 

 

 

 

 

 

 

 

 

 

C. (DeVilbiss LT Series) Unit operates, the Power light is on

1.

Air filter is blocked.

1.

Check the air filter. If the filter is dirty, wash it following the

 

 

when the power switch is “On.” Red Service Required light is

 

 

 

cleaning instructions on page 18.

 

 

illuminated.Audible alarm may be sounding.

2.

Exhaust is blocked.

2.

Check the exhaust area; make sure there is nothing restricting

 

 

 

 

 

 

the unit exhaust.

 

 

 

 

 

 

If the above remedies do not work, contact your

 

 

 

 

 

 

Sunrise Medical provider.

 

 

 

 

 

 

 

 

 

 

 

D. (DeVilbiss 3 & 5 Liter Series) Limited flow of oxygen,audible alarm is

1.

Blocked or defective cannula, catheter, face mask,

1

Detach cannula, catheter, or face mask. If proper flow is restored,

 

 

pulsing and Service Required light is flashing.

 

or oxygen tubing.

 

clean or replace if necessary. Disconnect the oxygen tubing at the

 

 

 

 

 

 

oxygen outlet. If proper flow is restored, check oxygen tubing for

 

 

 

 

 

 

obstructions or kinks. Replace if necessary.

 

 

 

2.

Blocked or defective humidifier bottle.

2.

Detach the humidifier from the oxygen outlet. If proper flow is

 

 

 

 

 

 

obtained, clean or replace humidifier.

 

 

 

3.

Flow meter set too low.

3.

Set flow meter to prescribed flow rate.

 

 

 

 

 

 

If the above remedies do not work, contact your

 

 

 

 

 

 

Sunrise Medical provider.

 

 

 

 

 

 

 

 

 

E. If any other problems occur with your oxygen concentrator.

 

 

1.

Turn your unit “Off.” Switch to your reserve oxygen system, and

 

 

 

 

 

 

contact your Sunrise Medical provider immediately.

 

 

 

 

 

 

 

 

 

MINOR TROUBLESHOOTING CHART - ONLY OSD Concentrators

 

 

Visual Indicator Lights

 

Possible Cause

 

Remedy

 

 

 

A. Both the green Normal Oxygen and the yellow Low Oxygen lights

1. OSD malfunction.

1. Contact your Sunrise Medical provider.

 

 

are either on or off.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B. Yellow Low Oxygen light is on or the yellow Low Oxygen light is on

1. Flow meter is not properly set.

1. Ensure the flow meter is properly set to the prescribed number.

 

 

and the intermittent audible signal is sounding.

2.

Air filter is blocked.

2.

Check the air filter. If the filter is dirty, wash it following the

 

 

 

 

 

 

cleaning instructions on page 18.

 

 

 

3. Exhaust is blocked.

3. Check the exhaust area; make sure there is nothing restricting the

 

 

 

 

 

 

unit exhaust. If the above remedies do not work, contact your

 

 

 

 

 

 

Sunrise Medical provider.

 

19

SP-505 Rev. B

Image 19
Contents Guide d’instructions Concentrateur ’oxygène DeVilbiss Inhaltsverzeichnis Table of ContentsIndice Table DES MatièresSound Level Specifications501DZ Oxygen PercentageEspecificaciones Organisme d’approbation CaractéristiquesPour les appareils équipés d’un Classe et type d’équipementAusrüstungsklasse und Typ Technische DatenEmpfohlener Maximaldurchfluß BetriebshöheSpecifiche Dichiarazione di Conformità Declaration of ConformityDéclaration deConformité KonformitätserklärungLesen Sie bitte die gesamte GEFAHR-RAUCHEN VerbotenCómo funciona su Concentradors de Oxígeno DeVilbiss Why Your Physician Prescribed Supplemental OxygenHow Your DeVilbiss Oxygen Concentrator Works Por qué su médico le pre- scribió un suplemento de oxígenoWichtige Teile Ihres O2 Konzentrators Important Parts of Your ConcentratorPartes IMPOR- Tantes DE SU Concentrador Pièces Importantes DE Votre Appareil ConcentrateurBack View AccessoriesAccessoires AccessoriBefore Operating Your Oxygen Concentrator Setting UP Your Oxygen CONCEN- TratorInstallation DE Votre Appareil Concentrateur D’OXYGÈNE Installazione DEL ConcentratoreConexión tubo de oxígeno con humidificación Oxygen Tubing Connection With HumidificationConnexion du tube d’oxy- gène à l’humidificateur Collegamento della tubazione di ossigeno con umidificazioneDevilbiss CONCEN- Tradors DE Oxígeno OSD SOLA- Mente Operating Your Devilbiss Oxygen ConcentratorDevilbiss Oxygen Concentrators with OSD only USO DE SU ConcentradorÜberprüfen Sie den Cuidado DE SU Devilbiss Concentradors DE Oxígeno Reserve Oxygen SystemCaring for Your Devilbiss Oxygen Concentrator Sistema DE Oxígeno DE ReservaCánula, Mascarilla y Tubos Air Filter and Oxygen Outlet ConnectorCannula, Mask, and Tubing Exterior CabinetMinor Troubleshooting Chart Minor TroubleshootingSymptom Possible Cause Remedy Tabla DE Detección Y Solución DE Problemas Menores Detección Y Solución DE Problemas MenoresSíntoma Causa posible Solución Tableau DE Dépannage Mineur Symptômes Cause Possible SolutionVoyants indicateurs visuels Cause Possible Solution Dépannage MineurZuständigen DeVilbiss Fachhändler verständigen Fehlersuche BEI Kleineren ProblemenDiagramm ZUR Fehlersuche BEI Kleineren Problemen Symptom Mögliche Ursache AbhilfeFornitore della Sunrise Medical Individuazione DI Problemi MinoriTabella DI Individuazione DI Problemi Minori Sintomo Possibile causa RimedioDeVilbiss Oxygen Concentrator Serial Number Check one Données sur le prestataire de ServicePhysician Information Oxygen use per dayPage Sunrise Medical SP-505 Rev. B

303, 501, 505 specifications

Sunrise Medical is a renowned name in the realm of mobility solutions, dedicated to enhancing the lives of individuals through innovative assistive technology. Among its prominent offerings are the Sunrise Medical 505, 501, and 303 models, each designed with a unique set of features, technologies, and characteristics tailored for different user needs.

The Sunrise Medical 505 is a lightweight and compact wheelchair designed for those who prioritize mobility and ease of use. One of its main features is the foldable frame, which allows for convenient transportation and storage. The 505 is equipped with a range of customization options, enabling users to select the seat width, depth, and backrest height that best suits their physique. This wheelchair is engineered with advanced maneuverability, thanks to its dual-axle design, which supports a smooth ride over various terrains. The 505's durable materials ensure longevity and resilience against daily wear and tear, while its intuitive joystick control allows users to navigate with precision and ease.

The Sunrise Medical 501 model stands out for its versatility and comfort. Designed with an emphasis on enhanced seating solutions, the 501 offers superior cushioning and support through its contoured seat and backrest. It features adjustable armrests and footrests, allowing users to modify the chair according to their comfort levels. Additionally, this model incorporates advanced technologies such as balancing systems that enhance stability and safety during movement. The 501 is particularly popular among users who require prolonged sitting periods, as it provides the necessary support to prevent pressure sores and ensure overall well-being.

On the other hand, the Sunrise Medical 303 embodies a blend of durability and advanced engineering. This model is often described as heavy-duty due to its reinforced structure that can withstand higher weight capacities while still maintaining stability. The 303 features an ergonomic design that prioritizes user comfort and ease of accessibility. Its innovative wheel design reduces rolling resistance, making it easier to navigate various surfaces. Furthermore, the 303 is equipped with puncture-proof tires, adding an extra layer of protection against flat tires and ensuring that users can depend on their mobility device for everyday use.

In summary, Sunrise Medical’s 505, 501, and 303 models represent a commitment to innovation and quality in wheelchair design. Each model offers distinct features and technologies tailored to meet the diverse needs of users, enhancing mobility, comfort, and independence in their daily lives. By prioritizing user-centered design, Sunrise Medical continues to make significant strides in the field of assistive technology, facilitating improved quality of life for individuals worldwide.