Sunrise Medical 505, 303 Specifications, 501DZ, Oxygen Percentage, Sound Level, Approval Body

Page 3

ENGLISH

SPECIFICATIONS

 

 

DEVILBISS 5-LITER SERIES

DEVILBISS 3-LITER SERIES

DEVILBISS LT SERIES

 

 

 

 

 

 

 

 

 

Catalog Number

505DZ/505DS

505KS/505KZ/505NS

303DZ/303DS/303DA

303IZ/303IS

501DZ

501IZ

 

 

 

 

 

 

 

 

 

Delivery Rate

1/2 to 5 LPM

1/2 to 5 LPM

1/2 to 3 LPM

1/2 to 3 LPM

1/2 to 5 LPM

1/2 to 5 LPM

 

 

 

 

 

 

 

 

 

Oxygen Percentage

1 LPM = 94% ± 3%

220-230V~, 50Hz

1/2-3 LPM = 93% ± 3%

220-240V~, 50Hz

1-4 LPM = 93% ± 3%

220-240V~, 50Hz

 

 

2-4 LPM = 95% ± 3%

1 LPM = 93% ± 3%

 

1/2-3 LPM = 93% ± 3%

5 LPM = 88% +3% -5%

1-3 LPM = 93% ± 3%

 

 

5 LPM = 93% ± 3%

2-4 LPM = 95% ± 3%

 

220-230V~, 60Hz

 

4 LPM = 90% ± 3%

 

 

 

5 LPM = 85% +6% -3%

 

1/2-3 LPM = 93% ± 3%

 

5 LPM = 80% +3% -5%

 

 

 

230V~, 60Hz

 

 

 

230V~, 60Hz

 

 

 

1 LPM = 92% ± 3%

 

 

 

1-3 LPM = 93% ± 3%

 

 

 

2-4 LPM = 95% ± 3%

 

 

 

4 LPM = 90% +5% -3 %

 

 

 

5 LPM = 92% ± 3%

 

 

 

5 LPM = 82% ± 3%

 

 

 

 

 

 

 

 

 

Maximum Recommended Flow

5 LPM

4 LPM

3 LPM

3 LPM

5 LPM

5 LPM

 

 

 

 

 

 

 

 

 

Sound Level

49 dBAAvg.

45 dBAAvg.

51 dBAAvg.

51 dBAAvg.

51 dBAAvg.

49 dBAAvg.

 

 

 

 

 

 

 

 

 

Outlet Pressure

8.5 psig (58.6 kPa)

8.5 psig (58.6 kPa)

8.5 psig (58.6 kPa)

8.5 psig (58.6 kPa)

8.5 psig (58.6 kPa)

8.5 psig (58.6 kPa)

 

 

 

 

 

 

 

 

 

Electrical Rating

115V~, 60Hz, 4.4Amp

220-230V~, 50Hz, 1.7 Amp

115V~, 60Hz, 2.1 Amp

220-240V~, 50Hz, 1.1 Amp

115V~, 60Hz, 5 Amp

220-230V~, 50Hz, 1.7 Amp

 

 

 

230V~, 60 Hz, 2.3 Amp*

 

220-230V~, 60 Hz, 1.2 Amp

 

230V~, 60 Hz, 1.7 Amp

 

 

 

 

 

 

 

 

 

Operating Voltage Range

95-132V~, 60Hz

187-253V~, 50Hz

95-132V~, 60Hz

187-264V~, 50Hz

95-132V~, 60Hz

187-264V~, 50Hz

 

 

 

195.5-253V~, 60 Hz*

 

187-253V~, 60 Hz

 

193-253V~, 60 Hz

 

 

 

 

 

 

 

 

 

Power Consumption

400 Watts Average

330 Watts Avg @230V, 50Hz

225 Watts Average

240 Watts Avg @230V, 50Hz

415 Watts Average

315 Watts Avg @230V, 50Hz

 

 

 

340 Watts Avg @230V, 60Hz

 

270 Watts Avg @230V, 60Hz

 

360 Watts Avg @230V, 60Hz

 

 

 

 

 

 

 

 

 

Weight

51 lbs. (23 kg.)

53 lbs. (24 kg.)

37 lbs. (17 kg.)

41 lbs. (19 kg.)

47 lbs. (21 kg.)

47 lbs. (21 kg.)

 

 

 

 

 

 

 

 

 

Dimensions

27.75”H x 16”W x 14”D

27.75”H x 16”W x 14”D

23”H x 13.87”W x 12”D

23”H x 13.87”W x 12”D

27.75”H x 16”W x 14”D

27.75”H x 16”W x 14”D

 

 

(70.5 x 40.6 x 35.6 cm)

(70.5 x 40.6 x 35.6 cm)

(58.4 x 35.2 x 30.5 cm)

(58.4 x 35.2 x 30.5 cm)

(70.5 x 40.6 x 35.6 cm)

(70.5 x 40.6 x 35.6 cm)

 

 

 

 

 

 

 

 

 

Pressure Relief Valve

35 psig ± 3 psig

35 psig ± 3 psig

25 psig ± 3 psig

25 psig ± 3 psig

35 psig ± 3 psig

35 psig ± 3 psig

 

 

(241 kPa ± 20.7 kPa)

(241 kPa ± 20.7 kPa)

(172 kPa ± 20.7 kPa)

(172 kPa ± 20.7 kPa)

(241 kPa ± 20.7 kPa)

(241 kPa ± 20.7 kPa)

 

 

 

 

 

 

 

 

 

Operating System

Pressure Sensing/

Pressure Sensing/

Pressure Sensing/

Pressure Sensing/

Timed/

Timed/

 

 

Pressure Swing

Pressure Swing

Pressure Vacuum

Pressure Vacuum

Pressure Swing

Pressure Swing

 

 

 

 

 

 

 

 

 

Operating Altitude

 

Up to 1500 meters (4,921 feet) above sea level without degradation of performance across the operating voltage range.

 

 

(tested at 21ºC only)

From 1500 meters (4,921 feet) to 4000 meters (13,123 feet), the end user should check with their Sunrise Medical provider for the suitability.

 

 

 

 

 

 

 

 

 

 

Operating Environment

 

5°C to 40°C, humidity range of 15% to 95%, non-condensing*

 

 

 

 

 

 

 

 

 

 

 

Recommended Operating

 

5°C to 40°C, humidity range of 15% to 95%, non-condensing*

 

5°C to 35°C, humidity range of 15% to 95%, non-condensing

Ambient Range (tested at sea level)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

For units equipped with

 

 

 

 

 

 

 

an OSD, the visible “low oxygen”

85%±2%

82%+1%-3%

85%±2%

85%±2%

Not Applicable

Not Applicable

 

indicator will activate at the

 

 

 

 

 

 

 

following levels

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Storage Conditions

 

-40°C to 70°C, humidity range of 10% to 100%, including condensation

 

 

 

Equipment Class and Type

 

Class II Equipment Double Insulated

Type B Applied Part

 

 

 

 

 

 

 

 

 

 

 

Approval Body and

CSA

TUV

CSA

TUV

CSA

CSA

 

Safety Standard

CAN/CSA-C22.2

IEC 601-1 + A1 + A2

CAN/CSA-C22.2

IEC 601-1

CAN/CSA-C22.2

CAN/CSA-C22.2

 

 

No. 601.1-M90

EN 60601 + A1 + A2

No. 601.1-M90

ISO 8359:1991

No. 601.1-M90

No. 601.1-M90

 

 

IEC 601-1

ISO8359: 1996

IEC 601-1

Complies with ISO 8359:1996

IEC 601-1

IEC 601-1

 

 

 

(505KZ excluded)

(303DA excluded)

(303IZ excluded)

 

 

 

 

 

 

 

 

 

 

 

EMC Compliance

IEC801-2 through 5

IEC 601-1-2

IEC801-2 through 5

IEC 601-1-2

IEC801-2 through 5

IEC 601-1-2

 

To:

CISPR 11/EN55011

 

CISPR 11/EN55011

 

CISPR 11/EN55011

 

 

 

 

 

 

 

 

 

Specifications subject to change without notice.

*505KS/505KZ/505NS = 25ºC maximum ambient for 60Hz rating.

3

SP-505 Rev. B

Image 3
Contents Guide d’instructions Concentrateur ’oxygène DeVilbiss Inhaltsverzeichnis Table of ContentsIndice Table DES MatièresSound Level Specifications501DZ Oxygen PercentageEspecificaciones Organisme d’approbation CaractéristiquesPour les appareils équipés d’un Classe et type d’équipementAusrüstungsklasse und Typ Technische DatenEmpfohlener Maximaldurchfluß BetriebshöheSpecifiche Dichiarazione di Conformità Declaration of ConformityDéclaration deConformité KonformitätserklärungLesen Sie bitte die gesamte GEFAHR-RAUCHEN VerbotenCómo funciona su Concentradors de Oxígeno DeVilbiss Why Your Physician Prescribed Supplemental OxygenHow Your DeVilbiss Oxygen Concentrator Works Por qué su médico le pre- scribió un suplemento de oxígenoWichtige Teile Ihres O2 Konzentrators Important Parts of Your ConcentratorPartes IMPOR- Tantes DE SU Concentrador Pièces Importantes DE Votre Appareil ConcentrateurBack View AccessoriesAccessoires AccessoriBefore Operating Your Oxygen Concentrator Setting UP Your Oxygen CONCEN- TratorInstallation DE Votre Appareil Concentrateur D’OXYGÈNE Installazione DEL ConcentratoreConexión tubo de oxígeno con humidificación Oxygen Tubing Connection With HumidificationConnexion du tube d’oxy- gène à l’humidificateur Collegamento della tubazione di ossigeno con umidificazioneDevilbiss CONCEN- Tradors DE Oxígeno OSD SOLA- Mente Operating Your Devilbiss Oxygen ConcentratorDevilbiss Oxygen Concentrators with OSD only USO DE SU ConcentradorÜberprüfen Sie den Cuidado DE SU Devilbiss Concentradors DE Oxígeno Reserve Oxygen SystemCaring for Your Devilbiss Oxygen Concentrator Sistema DE Oxígeno DE ReservaCánula, Mascarilla y Tubos Air Filter and Oxygen Outlet ConnectorCannula, Mask, and Tubing Exterior CabinetMinor Troubleshooting Minor Troubleshooting ChartSymptom Possible Cause Remedy Detección Y Solución DE Problemas Menores Tabla DE Detección Y Solución DE Problemas MenoresSíntoma Causa posible Solución Tableau DE Dépannage Mineur Symptômes Cause Possible SolutionVoyants indicateurs visuels Cause Possible Solution Dépannage MineurZuständigen DeVilbiss Fachhändler verständigen Fehlersuche BEI Kleineren ProblemenDiagramm ZUR Fehlersuche BEI Kleineren Problemen Symptom Mögliche Ursache AbhilfeFornitore della Sunrise Medical Individuazione DI Problemi MinoriTabella DI Individuazione DI Problemi Minori Sintomo Possibile causa RimedioDeVilbiss Oxygen Concentrator Serial Number Check one Données sur le prestataire de ServicePhysician Information Oxygen use per dayPage Sunrise Medical SP-505 Rev. B