Invacare TPO100 Power Input, External Power Supply Input, Sound Level, Altitude, Sensitivity

Page 15

SECTION 5—TYPICAL PRODUCT PARAMETERS

POWER INPUT:

24 VDC @ 7.5 Amp

 

or

 

11-16 VDC @ 10.0 Amp max

 

(12.6 VDC Nom.)

 

 

EXTERNAL POWER SUPPLY

 

INPUT:

 

AC POWER SUPPLY:

120 VAC, 50/60 Hz @ 2.5 amps

 

230 VAC, 50/60 Hz @ 1.25 amps

DC POWER SUPPLY:

11-16 VDC,

 

(12.6 VDC Nom. @ 10.0 Amp max)

 

 

SOUND LEVEL:

< 40 dBA weighted @ 2 LPM continuous and

 

all pulse settings (1-5)

 

 

ALTITUDE:

Up to 10,000 ft (3046 m) above sea level

 

Titration recommended for use above 10,000

 

ft (3046 m)

 

 

OXYGEN CONCENTRATION:*

87% to 95.6%, after initial warm-up period

*BASED ON AN ATMOSPHERIC

(approximately 5 minutes) at all flow rates

PRESSURE OF 14.7 PSI (101 KPA) AT

 

70°F (21°C) NOMINAL

 

 

 

CONSERVER TRIGGER

< 0.20 cmH2O max pressure drop

SENSITIVITY:

 

FACTORY SET - NO ADJUSTMENT

All settings - using 7 ft (2.1 m) cannula

PRESSURE ACTIVATED

 

 

 

CONSERVER TRIGGER DELAY:

35 mSec max

FACTORY SET - NO ADJUSTMENT

 

 

Nominal value - using 7 ft (2.1 m) cannula

 

 

CONSERVER BOLUS

Setting 1: 400 cc

DELIVERY:

Setting 2: 800 cc

FIXED MINUTE VOLUME

Setting 3: 1200 cc

 

Setting 4: 1600 cc

 

Setting 5: 2000 cc

 

Total cc’s ± 75 cc delivered per minute

 

 

CONSERVER BREATH RATE

Up to, and including, 35 BPM without

CAPACITY:

reduction of bolus minute volume

 

 

CONTINUOUS FLOW

0.5 through 3.0 LPM @ 0 psi

SETTINGS:

0.5 LPM increments

 

All settings are ± 0.2 LPM

 

(2.0 LPM max w/Ext DC Power Cable)

 

 

Part No 1156872

15

Invacare SOLO2™

Image 15
Contents Invacare SOLO2 Transportable Oxygen Concentrator Accessories Warning Table of Contents Special Notes Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information General Guidelines Maintenance Radio Frequency InterferencePolarized Plug Instruction Storage Packaging and HandlingUnpacking InspectionTechnical Description RED FeaturesNo AP/APG Typical Product ParametersSettings Power InputExternal Power Supply Input Conserver Trigger DelayBattery Recharge Time LPM Continuous DimensionsWeight Battery DurationRegulatory Listing Filters Operating InstructionsLocation Recommendations for Optimal PerformancePowering the Transportable Oxygen Concentrator Detail B DC Power Cable Detail C AC Power Adapter Detail a Battery ModuleInstalling the Battery Module Charging the BatteryBattery Time Management Prolonging Battery Life Checking Battery Charge LevelConnecting/Positioning the Nasal Cannula Detail aDetail B Remove cap from bottle Connecting the HumidifierDetail a Detail B Indicators and Audible AlarmControl/Adjustment Buttons LCD Displays Mode & Flow Adjustment Screen Typical Warning Screen Typical Alarm Screen Invacare SOLO2 Detail B LCD Display Icons Detail a Front PanelUsing the SOLO2 Transportable Oxygen Concentrator CLEANING, CARE, and Maintenance CabinetAir Intake Filter Cleaning the Humidifier Cleaning the CannulaTransportable Oxygen Concentrator Preventative Maintenance RecordTroubleshooting Troubleshooting GuideUnit is Beeping with no Lights or Display Symptom Probable Cause SolutionLCD Display Indicators Description SolutionsAcknowledgement Required Double audible beep every 15 sec Turn unit Off. Wait one minute, then Move unit to warmer, or cooler Troubleshooting Guide Ten audible beeps every 10 sec Ten audible beeps every 10 sec RED indicator Flashing Fast Continuous audible beep every half Troubleshooting Guide Make sure that car boat or motor Optional Accessories Limited Warranty Limited Warranty European Contacts European Contacts EC REP USA