SECTION
LCD DISPLAY: | INDICATORS: | |
|
| |
| Continuous audible beep every half | |
| second. | |
| RED indicator is On. | |
| GREEN indicator is Off. | |
|
| |
DESCRIPTION: | SOLUTIONS: | |
|
| |
| 1.Verify unit is in recommended | |
| environmental temperature. Refer to | |
| Typical Product Parameters on page 14. | |
The oxygen output purity has fallen to a | 2. | Clean intake filter and ensure both |
value below 73% (± 1%). | intake and exhaust are not blocked | |
| 3. | Turn unit Off, then On again to retry. |
| 4. | Change to another source of oxygen if |
| alarm continues. | |
|
|
|
ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart.
Call your provider and report the problem if alarm continues with next use.
LCD DISPLAY: | INDICATORS: | |
|
| |
| Continuous audible beep every half | |
| second. | |
| RED indicator is On. | |
| GREEN indicator is Off. | |
|
| |
DESCRIPTION: | SOLUTIONS: | |
|
| |
Abnormal system condition has been | 1. Turn unit Off, then On again to retry. | |
2. Change to another source of oxygen if | ||
detected | ||
alarm continues. | ||
| ||
|
|
ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart. Call your provider and report the problem if alarm continues.
Invacare SOLO2™ | 44 | Part No 1156872 |