SECTION
LCD DISPLAY: | INDICATORS: |
|
|
| Single audible beep every 15 sec |
| YELLOW indicator On. |
|
|
DESCRIPTION: | SOLUTIONS: |
|
|
| 1. Move unit to warmer, or cooler, |
(Only possible when External Power | surroundings. Allow unit to cool down to |
less than 95° F (35° C), or warm up to 50° | |
is applied) | F (10° C). Use AC or DC power. |
| |
Battery is too hot, or cold, to allow for the | 2. Remove battery and only use AC or DC |
battery to charge. | power. |
| |
| 3. Turn unit Off, then On again to retry. |
|
|
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens.
If the warning is issued and acknowledged a second time, the warning will be deactivated until the unit is turned off.
Call your provider and report the problem if the warning continues with next use.
LCD DISPLAY: | INDICATORS: | |||
|
|
|
| |
|
|
| Triple audible beep every 10 sec | |
|
|
| ||
|
|
| YELLOW indicator Flashing. | |
|
|
|
| |
DESCRIPTION: | SOLUTIONS: | |||
|
|
|
| |
External DC Power is lower than optimal. | 1. Make sure that car (boat or motor | |||
Excessive current draw could be present. | home, etc.) is running. | |||
2. Make sure DC Power Cable is firmly | ||||
Warning will continue every 3 minutes as | ||||
connected at both ends. | ||||
long as condition exists. | 3. Switch to external AC or battery power. | |||
|
|
| ||
|
|
|
|
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens
Invacare SOLO2™ | 40 | Part No 1156872 |