Invacare TPO100B user manual Move unit to warmer, or cooler

Page 40

SECTION 8—TROUBLESHOOTING GUIDE

LCD DISPLAY:

INDICATORS:

 

 

 

Single audible beep every 15 sec

 

YELLOW indicator On.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

1. Move unit to warmer, or cooler,

(Only possible when External Power

surroundings. Allow unit to cool down to

less than 95° F (35° C), or warm up to 50°

is applied)

F (10° C). Use AC or DC power.

 

Battery is too hot, or cold, to allow for the

2. Remove battery and only use AC or DC

battery to charge.

power.

 

 

3. Turn unit Off, then On again to retry.

 

 

ACKNOWLEDGEMENT REQUIRED:

Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens.

If the warning is issued and acknowledged a second time, the warning will be deactivated until the unit is turned off.

Call your provider and report the problem if the warning continues with next use.

LCD DISPLAY:

INDICATORS:

 

 

 

 

 

 

 

Triple audible beep every 10 sec

 

 

 

 

 

 

YELLOW indicator Flashing.

 

 

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

 

External DC Power is lower than optimal.

1. Make sure that car (boat or motor

Excessive current draw could be present.

home, etc.) is running.

2. Make sure DC Power Cable is firmly

Warning will continue every 3 minutes as

connected at both ends.

long as condition exists.

3. Switch to external AC or battery power.

 

 

 

 

 

 

 

ACKNOWLEDGEMENT REQUIRED:

Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens

Invacare SOLO2

40

Part No 1156872

Image 40
Contents Invacare SOLO2 Transportable Oxygen Concentrator Accessories Warning Table of Contents Special Notes Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information General Guidelines Radio Frequency Interference MaintenancePolarized Plug Instruction Packaging and Handling UnpackingInspection StorageTechnical Description Features REDTypical Product Parameters No AP/APGPower Input External Power Supply InputConserver Trigger Delay SettingsLPM Continuous Dimensions WeightBattery Duration Battery Recharge TimeRegulatory Listing Operating Instructions LocationRecommendations for Optimal Performance FiltersPowering the Transportable Oxygen Concentrator Detail a Battery Module Detail B DC Power Cable Detail C AC Power AdapterCharging the Battery Installing the Battery ModuleBattery Time Management Checking Battery Charge Level Prolonging Battery LifeDetail a Connecting/Positioning the Nasal CannulaDetail B Connecting the Humidifier Remove cap from bottleIndicators and Audible Alarm Detail a Detail BControl/Adjustment Buttons LCD Displays Mode & Flow Adjustment Screen Typical Warning Screen Typical Alarm Screen Invacare SOLO2 Detail a Front Panel Detail B LCD Display IconsUsing the SOLO2 Transportable Oxygen Concentrator Cabinet CLEANING, CARE, and MaintenanceAir Intake Filter Cleaning the Cannula Cleaning the HumidifierPreventative Maintenance Record Transportable Oxygen ConcentratorTroubleshooting Guide TroubleshootingSymptom Probable Cause Solution Unit is Beeping with no Lights or DisplayDescription Solutions LCD Display IndicatorsAcknowledgement Required Double audible beep every 15 sec Turn unit Off. Wait one minute, then Move unit to warmer, or cooler Troubleshooting Guide Ten audible beeps every 10 sec Ten audible beeps every 10 sec RED indicator Flashing Fast Continuous audible beep every half Troubleshooting Guide Make sure that car boat or motor Optional Accessories Limited Warranty Limited Warranty European Contacts European Contacts USA EC REP