Invacare TPO100B user manual Troubleshooting Guide

Page 45

SECTION 8—TROUBLESHOOTING GUIDE

 

 

LCD DISPLAY:

INDICATORS:

 

 

 

Continuous audible beep every half

 

second.

 

RED indicator is On.

 

GREEN indicator is Off.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

Abnormal compressor operation has been

1. Turn unit Off, then On again to retry.

detected.

2. Change to another source of oxygen if

alarm continues.

 

 

 

ACKNOWLEDGEMENT REQUIRED:

 

 

 

Press and hold the On/Off button to power down the unit and restart.

Call your provider and report the problem if alarm continues.

 

 

 

 

LCD DISPLAY:

INDICATORS:

 

 

 

Triple audible beep every 10 sec

 

YELLOW indicator Flashing.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

Oxygen sensor feedback is out-of-range.

1. Turn unit Off, then On again to retry.

Sensor failure is likely

 

 

 

ACKNOWLEDGEMENT REQUIRED:

 

 

 

Press Return / Highlight button. Display will either show additional warnings / alarms or

revert to normal operating screens.

 

If the alarm is issued and acknowledged a second time, the audible alarm will be

deactivated until the unit is turned off.

 

Call your provider and report the problem.

 

 

 

Part No 1156872

45

Invacare SOLO2™

Image 45
Contents Invacare SOLO2 Transportable Oxygen Concentrator Accessories Warning Table of Contents Special Notes Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information General Guidelines Maintenance Radio Frequency InterferencePolarized Plug Instruction Unpacking Packaging and HandlingInspection StorageTechnical Description RED FeaturesNo AP/APG Typical Product ParametersExternal Power Supply Input Power InputConserver Trigger Delay SettingsWeight LPM Continuous DimensionsBattery Duration Battery Recharge TimeRegulatory Listing Location Operating InstructionsRecommendations for Optimal Performance FiltersPowering the Transportable Oxygen Concentrator Detail B DC Power Cable Detail C AC Power Adapter Detail a Battery ModuleInstalling the Battery Module Charging the BatteryBattery Time Management Prolonging Battery Life Checking Battery Charge LevelConnecting/Positioning the Nasal Cannula Detail aDetail B Remove cap from bottle Connecting the HumidifierDetail a Detail B Indicators and Audible AlarmControl/Adjustment Buttons LCD Displays Mode & Flow Adjustment Screen Typical Warning Screen Typical Alarm Screen Invacare SOLO2 Detail B LCD Display Icons Detail a Front PanelUsing the SOLO2 Transportable Oxygen Concentrator CLEANING, CARE, and Maintenance CabinetAir Intake Filter Cleaning the Humidifier Cleaning the CannulaTransportable Oxygen Concentrator Preventative Maintenance RecordTroubleshooting Troubleshooting GuideUnit is Beeping with no Lights or Display Symptom Probable Cause SolutionLCD Display Indicators Description SolutionsAcknowledgement Required Double audible beep every 15 sec Turn unit Off. Wait one minute, then Move unit to warmer, or cooler Troubleshooting Guide Ten audible beeps every 10 sec Ten audible beeps every 10 sec RED indicator Flashing Fast Continuous audible beep every half Troubleshooting Guide Make sure that car boat or motor Optional Accessories Limited Warranty Limited Warranty European Contacts European Contacts EC REP USA