SECTION 5—TYPICAL PRODUCT PARAMETERS

EXTENDED TEMPERATURE

 

 

RANGE:

 

 

(USING BATTERY)

 

 

OPERATING TEMPERATURE:

95°F to 104° (35°C to 40°C)

 

 

 

 

Pulse Mode

Continuous Mode

 

 

 

 

Settings 1, 2 and 3 =

0.5 thru 1.5 LPM =

 

unlimited use

unlimited use

 

 

 

 

Setting 4 = 45

2.0 thru 2.5 LPM = 45

 

minutes

minutes

 

 

 

 

Setting 5 = 30

3.0 LPM =

 

minutes

30 minutes

 

 

 

Regulatory Listing

ETL certified complying with:

EN 55011: 1998

CISPR 11: 2003

IEC 60601-1: 2005

IEC 60601-1-2: 2.1 Ed.

IEC 61000-3-2: 2005

IEC 61000-3-3: 2005

UL 60601-1, 1st ed.

CSA 601.1 M90

ISO 8359

Part No 1156872

17

Invacare SOLO2

Image 17
Contents Invacare SOLO2 Transportable Oxygen Concentrator Accessories Warning Table of Contents Special Notes Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information General Guidelines Polarized Plug Instruction MaintenanceRadio Frequency Interference Unpacking Packaging and HandlingInspection StorageTechnical Description RED FeaturesNo AP/APG Typical Product ParametersExternal Power Supply Input Power InputConserver Trigger Delay SettingsWeight LPM Continuous DimensionsBattery Duration Battery Recharge TimeRegulatory Listing Location Operating InstructionsRecommendations for Optimal Performance FiltersPowering the Transportable Oxygen Concentrator Detail B DC Power Cable Detail C AC Power Adapter Detail a Battery ModuleBattery Time Management Installing the Battery ModuleCharging the Battery Prolonging Battery Life Checking Battery Charge LevelDetail B Connecting/Positioning the Nasal CannulaDetail a Remove cap from bottle Connecting the HumidifierDetail a Detail B Indicators and Audible AlarmControl/Adjustment Buttons LCD Displays Mode & Flow Adjustment Screen Typical Warning Screen Typical Alarm Screen Invacare SOLO2 Detail B LCD Display Icons Detail a Front PanelUsing the SOLO2 Transportable Oxygen Concentrator Air Intake Filter CLEANING, CARE, and MaintenanceCabinet Cleaning the Humidifier Cleaning the CannulaTransportable Oxygen Concentrator Preventative Maintenance RecordTroubleshooting Troubleshooting GuideUnit is Beeping with no Lights or Display Symptom Probable Cause SolutionAcknowledgement Required LCD Display IndicatorsDescription Solutions Double audible beep every 15 sec Turn unit Off. Wait one minute, then Move unit to warmer, or cooler Troubleshooting Guide Ten audible beeps every 10 sec Ten audible beeps every 10 sec RED indicator Flashing Fast Continuous audible beep every half Troubleshooting Guide Make sure that car boat or motor Optional Accessories Limited Warranty Limited Warranty European Contacts European Contacts EC REP USA