SECTION 8—TROUBLESHOOTING GUIDE

LCD DISPLAY:

INDICATORS:

 

 

 

Triple audible beep every 10 sec

 

YELLOW indicator Flashing

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

(Only possible with Continuous Flow

1.

Verify the cannula is not kinked.

Mode)

2.

Replace outlet filter, if necessary.

Output flow is more than 0.3 LPM below

3.

Turn unit Off. Wait one minute, then

the flow setting.

retry.

 

 

 

ACKNOWLEDGEMENT REQUIRED:

Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens. This warning will be turned off for a minimum of 15 minutes.

If the warning is issued and acknowledged a second time, the warning will be deactivated until the unit is turned off.

Call your provider and report the problem if the warning continues with next use.

LCD DISPLAY:

INDICATORS:

 

 

 

Single audible beep every 15 sec

 

YELLOW indicator On.

 

 

DESCRIPTION:

SOLUTIONS:

 

 

 

1. Verify unit is in recommended

 

environmental temperature. Refer to

The oxygen output purity has fallen to a

Typical Product Parameters on page 14.

2. Clean intake filter and ensure both

value between 73% and 85% (± 1%).

 

intake and exhaust are not blocked

 

3. Turn unit Off, then On again to retry.

 

 

ACKNOWLEDGEMENT REQUIRED:

Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens.

If the warning is issued and acknowledged a second time, the warning will be deactivated until the unit is turned off.

Call your provider and report the problem if the warning continues with next use.

Part No 1156872

39

Invacare SOLO2

Image 39
Contents Invacare SOLO2 Transportable Oxygen Concentrator Accessories Warning Table of Contents Special Notes Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information General Guidelines Maintenance Radio Frequency InterferencePolarized Plug Instruction Storage Packaging and HandlingUnpacking InspectionTechnical Description RED FeaturesNo AP/APG Typical Product ParametersSettings Power InputExternal Power Supply Input Conserver Trigger DelayBattery Recharge Time LPM Continuous DimensionsWeight Battery DurationRegulatory Listing Filters Operating InstructionsLocation Recommendations for Optimal PerformancePowering the Transportable Oxygen Concentrator Detail B DC Power Cable Detail C AC Power Adapter Detail a Battery ModuleInstalling the Battery Module Charging the BatteryBattery Time Management Prolonging Battery Life Checking Battery Charge LevelConnecting/Positioning the Nasal Cannula Detail aDetail B Remove cap from bottle Connecting the HumidifierDetail a Detail B Indicators and Audible AlarmControl/Adjustment Buttons LCD Displays Mode & Flow Adjustment Screen Typical Warning Screen Typical Alarm Screen Invacare SOLO2 Detail B LCD Display Icons Detail a Front PanelUsing the SOLO2 Transportable Oxygen Concentrator CLEANING, CARE, and Maintenance CabinetAir Intake Filter Cleaning the Humidifier Cleaning the CannulaTransportable Oxygen Concentrator Preventative Maintenance RecordTroubleshooting Troubleshooting GuideUnit is Beeping with no Lights or Display Symptom Probable Cause SolutionLCD Display Indicators Description SolutionsAcknowledgement Required Double audible beep every 15 sec Turn unit Off. Wait one minute, then Move unit to warmer, or cooler Troubleshooting Guide Ten audible beeps every 10 sec Ten audible beeps every 10 sec RED indicator Flashing Fast Continuous audible beep every half Troubleshooting Guide Make sure that car boat or motor Optional Accessories Limited Warranty Limited Warranty European Contacts European Contacts EC REP USA