SECTION
Warnings and Alarms
NOTE: In all cases, pressing and holding the On/Off button for 2‐3 seconds will turn off and reset the unit.
LCD DISPLAY: | INDICATORS: | |
|
| |
| Triple audible beep every 10 sec | |
| YELLOW indicator Flashing. | |
|
| |
DESCRIPTION: | SOLUTIONS: | |
|
| |
(Only possible with Pulse Mode) | 1. Verify the cannula is connected, not | |
Transportable oxygen concentrator has | kinked, properly positioned and you are | |
breathing through your nose. | ||
not detected a breath for a predetermined | ||
time period. |
| |
|
| |
ACKNOWLEDGEMENT REQUIRED: |
| |
|
|
Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens. This warning will be turned off for a minimum of 5 minutes.
LCD DISPLAY: | INDICATORS: |
|
|
| Single audible beep every 15 sec |
| YELLOW indicator On. |
|
|
DESCRIPTION: | SOLUTIONS: |
|
|
(Only possible with Pulse Mode) | 1. Immediately reduce your activity level to |
Your breathing rate has begun to exceed | get your breath rate to slow down. |
| |
the capacity of the transportable oxygen | 2. Temporarily switch to Continuous Flow |
concentrator. | Mode. |
|
|
ACKNOWLEDGEMENT REQUIRED: |
|
|
|
A sufficient reduction of your breath rate will automatically turn off the warning.
Alternately, press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens. This warning will be turned off for a minimum of 5 minutes.
Part No 1156872 | 37 | Invacare SOLO2™ |