SECTION
NO CHARGING SYMBOL SHOWN ON STANDBY SCREEN
SYMPTOM | PROBABLE CAUSE | SOLUTION | |
With the unit turned | Power source is no good, or | Try another power outlet | |
Off and AC or DC | there is a loose connection. | and check connections to | |
power plugged into |
| unit. | |
the unit, no charging |
|
| |
Battery is outside the allowed | Allow unit to cool down to | ||
icon is displayed. | |||
temperature range for charging. | less than 95° F | ||
| |||
|
| (35° C), or warm up to 50° F | |
|
| (10° C). | |
|
|
| |
| Battery module is not fully | Ensure the battery module is | |
| seated. | completely inserted. | |
|
|
| |
| Standby screen is not being | Press the On/Off button to | |
| displayed. | display the Standby screen | |
|
| when using DC power. | |
|
|
| |
| |||
UNIT IS BEEPING WITH NO LIGHTS OR DISPLAY | |||
|
|
| |
SYMPTOM | PROBABLE CAUSE | SOLUTION | |
All power sources are | Battery pack has been removed | Install any battery pack | |
missing. | and there is no external power | available. | |
| While operating without a | Beeping will last | |
| battery, the external power is | minutes if no power source | |
| lost. | is provided. | |
|
|
|
Invacare SOLO2™ | 36 | Part No 1156872 |