SECTION
LCD DISPLAY: | INDICATORS: | |
|
| |
| Double audible beep every 15 sec | |
| YELLOW indicator On. | |
|
| |
DESCRIPTION: | SOLUTIONS: | |
|
| |
Remaining battery capacity has fallen to | 1. Connect transportable oxygen | |
concentrator to either the AC or DC | ||
25%. Battery requires charging. | ||
power, or use an alternate charged battery | ||
Battery gauge is empty | ||
module. | ||
| ||
|
|
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or revert to normal operating screens.
LCD DISPLAY: | INDICATORS: | ||
|
| ||
| Single audible beep every 15 sec | ||
| YELLOW indicator On. | ||
|
| ||
DESCRIPTION: | SOLUTIONS: | ||
|
|
| |
Unit is too hot, or too cold, to allow it to | 1. | Move unit to warmer, or cooler, | |
surroundings. Allow unit to cool down to | |||
operate. | less than 95° F (35° C), or warm up to 50° | ||
| |||
Unit will run internal fans to help lower, or | F (10° C). Use AC or DC power. | ||
raise, internal temperature. | 2. | Clean intake filter. | |
| |||
Fans will turn off after 10 minutes | 3. | Use | |
regardless of temperature. | |||
4. | Turn unit Off, then On again to retry. | ||
| |||
|
|
|
ACKNOWLEDGEMENT REQUIRED:
While the unit is warming / cooling the alarm screen will remain active. The unit will turn completely off (battery mode) or return to the standby screen (AC & DC mode), whenever the desired temperature is reached or 10 minutes of cooling has expired.
Invacare SOLO2™ | 38 | Part No 1156872 |