Invacare TPO100B user manual Control/Adjustment Buttons

Page 26

SECTION 6—OPERATING INSTRUCTIONS

Breath Detect Indication

(Pulse Mode Only) The GREEN indicator light will blink off each time an inhalation is detected and the transportable oxygen concentrator outputs a measured pulse of oxygen.

Audible Alarm

The transportable oxygen concentrator is equipped with an audible alarm. The audible alarm will activated in conjunction with Indicators and LCD Display to annunciate Alarms and Warnings.

The transportable oxygen concentrator will also issue a steady beep when all power sources are removed. The beeping will stop after 10‐20 minutes or once a power source is connected.

The transportable oxygen concentrator will briefly turn on all three lights and the LCD display and check for stuck buttons upon power up. It will then output a brief audible beep to indicate the unit is operational.

Control/Adjustment Buttons

There are four buttons located surrounding the LCD display. They are used to turn the unit on/off, to proceed to the next screen, to select and highlight adjustable settings and to adjust those settings.

On/Off Button

To turn the transportable oxygen concentrator on or off, press and hold the On/Off button for a period of 2‐3 seconds, not to exceed 5 seconds. Releasing the button too soon will not allow the unit to do either. If the button is held too long when turning on the unit, the Stuck Button alarm could be activated.

Mode/Return/Highlight Button

Pressing button for 1‐2 seconds in the Standby screen (FIGURE 6.5 on page 27) will switch to Normal Operating screen (FIGURE 6.6 on page 27).

Pressing this button for 1 second in the Normal Operating screen will toggle between the continuous and pulse modes.

Pressing this button for 2‐3 seconds in the Normal Operating screen will display the LCD Adjustment screen. In the LCD Adjustment screen pressing this button for 1 second will switch columns on the screen. To save changes and exit the LCD adjustment screen the highlight bar needs to be positioned in the left column, then press this button for 2‐3 seconds to return to the Normal Operating screen. If the screen switches without pressing and holding the button for 2‐3 seconds the changes made in the LCD Adjustment screen will not be saved.

Increase and Decrease Buttons

When the highlight bar is on the setting name, these buttons are used to select the next setting in the list. When the highlight bar is on the setting value, these buttons will change the value to the next value available.

Invacare SOLO2

26

Part No 1156872

Image 26
Contents Invacare SOLO2 Transportable Oxygen Concentrator Accessories Warning Table of Contents Special Notes Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information General Guidelines Polarized Plug Instruction MaintenanceRadio Frequency Interference Inspection Packaging and HandlingUnpacking StorageTechnical Description Features REDTypical Product Parameters No AP/APGConserver Trigger Delay Power InputExternal Power Supply Input SettingsBattery Duration LPM Continuous DimensionsWeight Battery Recharge TimeRegulatory Listing Recommendations for Optimal Performance Operating InstructionsLocation FiltersPowering the Transportable Oxygen Concentrator Detail a Battery Module Detail B DC Power Cable Detail C AC Power AdapterBattery Time Management Installing the Battery ModuleCharging the Battery Checking Battery Charge Level Prolonging Battery LifeDetail B Connecting/Positioning the Nasal CannulaDetail a Connecting the Humidifier Remove cap from bottleIndicators and Audible Alarm Detail a Detail BControl/Adjustment Buttons LCD Displays Mode & Flow Adjustment Screen Typical Warning Screen Typical Alarm Screen Invacare SOLO2 Detail a Front Panel Detail B LCD Display IconsUsing the SOLO2 Transportable Oxygen Concentrator Air Intake Filter CLEANING, CARE, and MaintenanceCabinet Cleaning the Cannula Cleaning the HumidifierPreventative Maintenance Record Transportable Oxygen ConcentratorTroubleshooting Guide TroubleshootingSymptom Probable Cause Solution Unit is Beeping with no Lights or DisplayAcknowledgement Required LCD Display IndicatorsDescription Solutions Double audible beep every 15 sec Turn unit Off. Wait one minute, then Move unit to warmer, or cooler Troubleshooting Guide Ten audible beeps every 10 sec Ten audible beeps every 10 sec RED indicator Flashing Fast Continuous audible beep every half Troubleshooting Guide Make sure that car boat or motor Optional Accessories Limited Warranty Limited Warranty European Contacts European Contacts USA EC REP