Invacare TPO100B user manual Troubleshooting Guide

Page 35

SECTION 8—TROUBLESHOOTING GUIDE

SECTION 8—TROUBLESHOOTING GUIDE

NOTE: In all cases, if the situation continues, change to another source of oxygen and contact your equipment provider.

Troubleshooting

TRANSPORTABLE OXYGEN CONCENTRATOR DOES NOT TURN ON, OR DOES NOT STAY ON

SYMPTOM

PROBABLE CAUSE

SOLUTION

 

Concentrator does

On/Off button was not held

Try to power up the unit

 

not operate when

down long enough.

again while continuing to

On/Off button

 

press the On/Off button until

pressed

 

the front panel lights and

 

 

LCD begin to illuminate. This

OR

 

takes typically 2-3 seconds.

 

 

 

 

LCD and front panel

Battery is discharged (or

Ensure the battery module is

lights DO NOT

improperly connected to the

completely inserted.

illuminate upon

unit).

Connect transportable

start-up

 

oxygen concentrator to

 

 

either AC or DC power and

OR

 

retry.

 

 

Battery pack has performed a

Wait one minute and retry.

All LCD and front

self-protect shutdown when

Change to an alternate

panel lights go dark.

there was no External Power.

battery pack.

 

 

 

 

 

 

GREEN INDICATOR DOES NOT BLINK WITH EACH BREATH

 

 

 

 

 

SYMPTOM

PROBABLE CAUSE

SOLUTION

 

No Breath Detect

Unit is set for Continuous Flow

No action required. Breath

 

indicator

operation.

Detect Indicator is for Pulse

 

 

 

Mode only

 

 

 

 

 

 

NO EXTERNAL POWER SYMBOL SHOWN ON STANDBY SCREEN

 

 

 

 

 

SYMPTOM

PROBABLE CAUSE

SOLUTION

 

With AC or DC

Power source is no good, or

Try another power outlet

 

power plugged into

there is a loose connection.

and check connections to

 

the unit, no external

 

unit.

 

power icon is

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

Part No 1156872

35

Invacare SOLO2

Image 35
Contents Invacare SOLO2 Transportable Oxygen Concentrator Accessories Warning Table of Contents Special Notes Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information General Guidelines Polarized Plug Instruction MaintenanceRadio Frequency Interference Storage Packaging and HandlingUnpacking InspectionTechnical Description RED FeaturesNo AP/APG Typical Product ParametersSettings Power InputExternal Power Supply Input Conserver Trigger DelayBattery Recharge Time LPM Continuous DimensionsWeight Battery DurationRegulatory Listing Filters Operating InstructionsLocation Recommendations for Optimal PerformancePowering the Transportable Oxygen Concentrator Detail B DC Power Cable Detail C AC Power Adapter Detail a Battery ModuleBattery Time Management Installing the Battery ModuleCharging the Battery Prolonging Battery Life Checking Battery Charge LevelDetail B Connecting/Positioning the Nasal CannulaDetail a Remove cap from bottle Connecting the HumidifierDetail a Detail B Indicators and Audible AlarmControl/Adjustment Buttons LCD Displays Mode & Flow Adjustment Screen Typical Warning Screen Typical Alarm Screen Invacare SOLO2 Detail B LCD Display Icons Detail a Front PanelUsing the SOLO2 Transportable Oxygen Concentrator Air Intake Filter CLEANING, CARE, and MaintenanceCabinet Cleaning the Humidifier Cleaning the CannulaTransportable Oxygen Concentrator Preventative Maintenance RecordTroubleshooting Troubleshooting GuideUnit is Beeping with no Lights or Display Symptom Probable Cause SolutionAcknowledgement Required LCD Display IndicatorsDescription Solutions Double audible beep every 15 sec Turn unit Off. Wait one minute, then Move unit to warmer, or cooler Troubleshooting Guide Ten audible beeps every 10 sec Ten audible beeps every 10 sec RED indicator Flashing Fast Continuous audible beep every half Troubleshooting Guide Make sure that car boat or motor Optional Accessories Limited Warranty Limited Warranty European Contacts European Contacts EC REP USA