SECTION 6—OPERATING INSTRUCTIONS

Connecting the Humidifier

(if so prescribed and only with Continuous Flow Mode)

WARNING

DO NOT use a humidifier while in Pulse Mode. DO NOT overfill humidifier.

DO NOT reverse the oxygen input and output connections. Water from the humidifier bottle will travel through the cannula back to the patient.

NOTE: For this procedure, refer to FIGURE 6.3 and FIGURE 6.4 on page 25.

NOTE: Connect the humidifier only if prescribed. Use the humidifier only in Continuous Flow Mode.

1.Remove cap from bottle.

2.Fill humidifier with distilled water to the level indicated by the manufacturer. Replace the humidifier cap and securely tighten.

Humidifier Bottle

Humidifier Bottle

w/o Cap

with Cap

 

FIGURE 6.3 Filling the Humidifier

3.Attach the humidifier bottle adapter to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. See DETAIL “A” in FIGURE 6.4.

4.Place the humidifier bottle/adapter assembly in the outside pocket of the accessory bag. Attach the accessory bag to the pull cart as shown in FIGURE 6.4.

5.Rotate the bottle within the pocket so that the humidifier adapter hose is away from the body of the transportable oxygen concentrator.

6.Attach oxygen tubing from the humidifier bottle/adapter assembly to oxygen outlet connector on the concentrator. See DETAIL “B” in FIGURE 6.4.

7.Attach the cannula/patient supply tubing to the humidifier bottle outlet. See DETAIL “B” in FIGURE 6.4.

8.After assembly, ensure that oxygen is flowing through the cannula.

Invacare SOLO2

24

Part No 1156872

Image 24
Contents Invacare SOLO2 Transportable Oxygen Concentrator Accessories Warning Table of Contents Special Notes Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information General Guidelines Maintenance Radio Frequency InterferencePolarized Plug Instruction Packaging and Handling UnpackingInspection StorageTechnical Description Features REDTypical Product Parameters No AP/APGPower Input External Power Supply InputConserver Trigger Delay SettingsLPM Continuous Dimensions WeightBattery Duration Battery Recharge TimeRegulatory Listing Operating Instructions LocationRecommendations for Optimal Performance FiltersPowering the Transportable Oxygen Concentrator Detail a Battery Module Detail B DC Power Cable Detail C AC Power AdapterInstalling the Battery Module Charging the BatteryBattery Time Management Checking Battery Charge Level Prolonging Battery LifeConnecting/Positioning the Nasal Cannula Detail aDetail B Connecting the Humidifier Remove cap from bottleIndicators and Audible Alarm Detail a Detail BControl/Adjustment Buttons LCD Displays Mode & Flow Adjustment Screen Typical Warning Screen Typical Alarm Screen Invacare SOLO2 Detail a Front Panel Detail B LCD Display IconsUsing the SOLO2 Transportable Oxygen Concentrator CLEANING, CARE, and Maintenance CabinetAir Intake Filter Cleaning the Cannula Cleaning the HumidifierPreventative Maintenance Record Transportable Oxygen ConcentratorTroubleshooting Guide TroubleshootingSymptom Probable Cause Solution Unit is Beeping with no Lights or DisplayLCD Display Indicators Description SolutionsAcknowledgement Required Double audible beep every 15 sec Turn unit Off. Wait one minute, then Move unit to warmer, or cooler Troubleshooting Guide Ten audible beeps every 10 sec Ten audible beeps every 10 sec RED indicator Flashing Fast Continuous audible beep every half Troubleshooting Guide Make sure that car boat or motor Optional Accessories Limited Warranty Limited Warranty European Contacts European Contacts USA EC REP