MODEL | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
5.4 CHOOSING DISPLAY OPTIONS (CONTINUED)
Temp units: °C
Output 1: mA
Output 2: mA
ExitEdit
Language: English
Display left: I
Display right: Out 2
ExitEdit
Display contrast: 40
Timeout: On
Timeout value: 10 min
ExitEdit
7.Set the remainder of the display parameters. Use the and keys to choose the desired parameter. Then press Edit (F4). Use the key to move the cursor to the desired selection. Press Save (F4) to store.
| Temp units | °C or °F |
|
|
|
|
|
| Output 1 | mA or % of full scale |
|
|
|
|
|
| Output 2 | mA or % of full scale |
|
|
|
|
|
| Language | English, Français, Español, Deutsch, Italiano |
|
|
|
|
|
| Display left | sensor current (I), alarm 1 setpoint (no units), |
|
|
| alarm 3 setpoint (no units), or blank |
|
|
|
|
|
| Display right | sensor current (I), alarm 2 setpoint (no units), |
|
|
| alarm 3 setpoint (no units), output 2, or blank |
|
|
|
|
|
| Display Contrast | 00 |
|
|
| changes as the number changes |
|
|
|
|
|
| Timeout | Timeout returns the display from any other screen |
|
|
| to the main display if no key is pressed before the |
|
| Timeout value |
| |
| timeout value is exceeded. |
| |
|
|
| |
|
|
|
|
| Polling address | Identifies analyzer in |
|
|
|
|
|
SECURITY CAUTION
The analyzer uses the timeout value to activate security. Once the analyzer is unlocked by entering a security code, security will not
28