Emerson 51-DO-03, 51-DO-04 instruction manual Noise rejection Main sensor cal Security

Page 45

MODEL DO-03/04

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.11SECURITY

1.The analyzer can be programmed to require a password for access to menus. There are three levels: Level 1: A level 1 user can

1.Zero and calibrate the oxygen sensor

2.Calibrate the barometric pressure sensor

3.Change temperature compensation from automatic to manual and enter a manual compensation temperature

4.View diagnostic variables.

Level 2: A level 2 user can

1.Do everything a level 1 user can do

2.Change alarm setpoints

3.Rerange the 4-20 mA outputs

4.Manually test both outputs and all four alarm relays.

Level 3: A level 3 user has access to every menu item. Only a level 3 user can change passwords.

A person with no password can only view the main display.

Noise rejection Main sensor cal Security

2.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).

3.Move the cursor to "Configure" and press Enter (F4).

4.Move the cursor to "Security" and press Enter (F4).

Exit

Enter

 

 

5. Use the ￿ and ￿ keys to move through the list of items. To enter a password,

 

 

press Edit (F4). Use the arrow keys to enter a three-digit password. Press

 

 

Save (F4) to store the value.

Lock all

: 0

Lock all: Until the user enters the "lock all" password, all he can do is view

Lock program: 0

the main display. Entering the "lock all" password allows the user access to

 

 

Lock config: 0

all Level 1 functions.

Lock program: Entering the "lock program" password allows the user

 

 

access to all Level 2 functions.

Exit

Edit

 

 

Lock config: Entering the "lock config" password allows the user access to

 

 

all Level 3 functions.

 

 

The analyzer will accept a higher level security code at a lower level securi-

 

 

ty gate. For example, the analyzer will accept a level 2 password at a level 1

 

 

gate.

NOTES:

a.A code of 000 disables security for that level.

b.The security feature will not activate until after the timeout period has passed with no key presses.

c.A hold condition will indefinitely prolong the timeout period.

d.Security will activate immediately if power is removed and then restored.

e.To recall a forgotten code, press and hold F4 for five seconds when the security screen appears. The code for that level will appear.

39

Image 45
Contents Model DO-03/04 Dissolved Oxygen Measurement System with Air Blast CleanerEssential Instructions Page Section Title Model DO-03/04 Measuring SystemList of Tables 10.011.0 List of Figures Section TitleIii Features and Benefits Section SpecificationsRepeatability ±0.05 ppm 25C PowerRFI/EMI EN-61326 Output Accuracy ± 0.05 mAMaximum continuous pressure 50 psig 446 kPa abs Maximum distance between compressor and sensorMaterials of Construction Analyzer dimensions Enclosure dimensions Power required selection Description9240048-00 Tag, stainless steel, specify marking Installation Section InstallationUnpacking and Inspection Model or part number DescriptionInstalling the air compressor enclosure Installing the handrail mounting assemblyAssembling the pipe and pipe clamp Installing the sensor in the washer head assembly Attaching the pipe boom to the handrail mounting assemblyAssembling the washer head and pipe boom Attaching the air compressor enclosure to the handrail Installing the sensor in the washer head assembly Section Wiring POWER, ALARM, and Output WiringGeneral 454EPH02 Section Section Sensor Wiring Wiring the Oxygen SensorGeneral Section Display and Operation Alarm StatusDisplay KEY Functions and ControlMenu Tree for the 54eA Analyzer Section Software ConfigurationOn following Main Menu Calibrate See Diagnostic Variable see Choices Factory Settings Setpoints Program Settings ListChoices Factory Settings Configure Choices Factory Settings Alarm 1 setpoint Alarm setpointsAlarm 2 setpoint Alarm 3 setpointRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mATest alarm 1 Open Simulated testsTest output Display Outputs AlarmsMeasure Oxygen Meas units ppmDisplay contrast Timeout On Temp units C Output 1 mA Output 2 mALanguage English Display left Display right Out Outputs Output 1 controlOutput Measurement Output 1 Setup Output 1 ControlRange 4-20 mA Dampen 0 sec Hold Last Value Alarm 1 control ConfigureAlarm 1 setup Alarm 2 control Changing Alarm ParametersAlarm 1 setup Alarm 1 controlAlarm Low Activate ProcessFeed limit Disable Alarm FaultTimeout 3600 sec Alarm 4 setupInterval timer Alarm 4 setup Feed limit timerTimer Disable Timer Time activated Interval 24.0 hr Interval timer setupTemp comp Auto Temperature Compensation and Temperature UnitsTemp units C Temp comp Manual Temp units CNoise Reduction 60 HzBar meas Auto Bar units mm HgBarometric Pressure Salinity 0.0 o/oo Main sensor calSecurity Noise rejection Main sensor cal SecurityAnalyzer Mode Priority Condition DefinitionsAction Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CSensor Current as a Function Dissolved Oxygen Concentration Section Calibration Dissolved OxygenZeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrate main sensor Zero main sensor Adjust temperatureStandardize Calibrating Barometric PressureTrimming the Outputs Section Calibration Current OutputsReplacement Parts Section MaintenanceCleaning the membrane Replacing electrolyte solution and membraneSection Spare Parts Power FuseAIR Blast Sensor Washer Replacement Parts Location Part Number DescriptionTroubleshooting When a Fault Message is Showing Section TroubleshootingFault message Explanation See Section OverviewHigh input current Temperature error low or highCheck sensor zero Sense line openTroubleshooting When no Fault Message is Showing Oxygen Problem See SectionZero current is too high Possible error warning during in-process calibration Zero reading Is unstableBarometric pressure reading is too high or too low Readings drift Process readings are erraticSensor does not respond to changes in oxygen level Oxygen readings are too lowProblem Action Troubleshooting not Related to Measurement ProblemsSimulating Inputs Dissolved Oxygen Sensor Polarizing Voltage Resistance Expected currentSimulating temperature Simulating TemperatureGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestWarranty Specifications subject to change without notice