MODEL | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
TABLE | (continued) |
|
|
|
|
ITEM | CHOICES | FACTORY SETTINGS |
CONFIGURE |
|
|
B Outputs (Section 5.5) |
|
|
1. Output 1 |
|
|
a. Measurement | Oxygen or temperature | Oxygen |
b. Control | Normal | Normal |
2. Output 1 Setup |
|
|
a. Current | ||
b. Dampening | 0 sec | |
c. Hold mode | Hold last value or go to fixed value | Hold last value |
d. Fixed hold value | 21 mA | |
e. Fault value | 22 mA | |
3. Output 2 |
|
|
a. Measurement | Oxygen or temperature | Temperature |
b. Control | Normal | Normal |
4. Output 2 Setup | See output 1 | See output 1 |
5. Hold feature | Enable, disable, or 20 min timeout | Disable |
C. Alarms (Section 5.6) |
|
|
1. Alarm 1 |
|
|
a. Activation method | Oxygen or temperature | Oxygen |
b. Control mode | Normal | Normal |
2. Alarm 1 setup |
|
|
a. Configuration | Low, high, or off | High |
b. Hysteresis |
|
|
if oxygen (ppm) | 0 to 20 ppm | 0 ppm |
if oxygen (ppb) | 0 to 999 ppb | 0 ppb |
if oxygen (% saturation) | 0 to 200% | 0% |
if temperature | 0 to 10°C | 0.1°C |
c. Delay time | 0 sec | |
d. Relay fault | none, open, closed | None |
3. Alarm 2 |
|
|
a. Activation method | Oxygen or temperature | Oxygen |
b. Control mode | Normal | Normal |
4. Alarm 2 setup |
|
|
a. Configuration | Low, high, or off | Low |
Rest of alarm 2 setup is the same as alarm 1
5.Alarm 3 setup is the same as alarm 1
6.Alarm 4
Alarm | Fault or off | Fault |
7. Feed limit timer |
|
|
a. Feed limit | Disable, alarm 1, alarm 2, or alarm 3 | Disable |
b. Timeout value | 0 to 10,800 sec | 600 sec |
8. Interval timer |
|
|
a. Select alarm | Disable, alarm 1, alarm 2, or alarm 3 | Disable |
b. Interval time | 0 to 999.9 hr | 24.0 hr |
c. Repeats | 1 to 60 | 1 |
d. On time | 0 to 2999 sec | 120 sec |
e. Off time | 0 to 2999 sec | 1 sec |
f. Recovery time | 0 to 999 sec | 600 sec |
22 | Continued on the following page |
|