Emerson 51-DO-03, 51-DO-04 instruction manual Adjust temp + 025.1 C, Temperature Calibration

Page 48

MODEL DO-03/04

SECTION 6.0

 

CALIBRATION - TEMPERATURE

6.2 TEMPERATURE CALIBRATION

Place the sensor and a standard thermometer in the process liquid.

1.Check the analyzer temperature reading (main display) to make sure the sensor has stabilized. Compare the analyzer temperature with the standard thermometer. The readings should differ by at most 1°C. If the read- ings differ by a greater amount, refer to Section 15.3. Go to the next step if the reading requires adjustment.

2.From the main display, press any key. With the cursor on “Calibrate,” press Enter (F4).

NOTE

Zero main sensor Adjust temperature Output trim

ExitEnter

25.1 °C

Adjust temp : 25.1 °C

ExitEdit

25.1 °C

Adjust temp : + 025.1 °C

If Hold was enabled in Section 5.5, the hold screen will appear. To activate hold, refer to Section 5.5, step 9.

3.Press the ￿ key twice to move the cursor to “Adjust temperature,” then press Enter (F4).

4.Press Edit (F4) to adjust the temperature. The screen below will then appear. Using the arrow keys, enter the correct temperature and press Save (F4). The controller will enter the value in memory. To abort the change, press Esc (F3). Press Exit (F1) three times for the main display.

NOTE

If Hold was activated during calibration, “Hold Mode Activated” will continue to flash in the main display. Return the sensor to normal and deactivate Hold. Refer to Section 5.5, step 9.

Esc Save

42

Image 48
Contents Dissolved Oxygen Measurement System with Air Blast Cleaner Model DO-03/04Essential Instructions Page Model DO-03/04 Measuring System Section TitleList of Tables 10.011.0 List of Figures Section TitleIii Section Specifications Features and BenefitsPower Repeatability ±0.05 ppm 25CRFI/EMI EN-61326 Output Accuracy ± 0.05 mAMaximum continuous pressure 50 psig 446 kPa abs Maximum distance between compressor and sensorMaterials of Construction Analyzer dimensions Enclosure dimensions Power required selection Description9240048-00 Tag, stainless steel, specify marking Section Installation InstallationUnpacking and Inspection Model or part number DescriptionInstalling the air compressor enclosure Installing the handrail mounting assemblyAssembling the pipe and pipe clamp Installing the sensor in the washer head assembly Attaching the pipe boom to the handrail mounting assemblyAssembling the washer head and pipe boom Attaching the air compressor enclosure to the handrail Installing the sensor in the washer head assembly Section Wiring POWER, ALARM, and Output WiringGeneral 454EPH02 Section Section Sensor Wiring Wiring the Oxygen SensorGeneral Alarm Status Section Display and OperationDisplay KEY Functions and ControlSection Software Configuration Menu Tree for the 54eA AnalyzerOn following Main Menu Calibrate See Diagnostic Variable see Program Settings List Choices Factory Settings SetpointsChoices Factory Settings Configure Choices Factory Settings Alarm setpoints Alarm 1 setpointAlarm 2 setpoint Alarm 3 setpointMA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mA Ranging the OutputsTest alarm 1 Open Simulated testsTest output Outputs Alarms DisplayMeasure Oxygen Meas units ppmDisplay contrast Timeout On Temp units C Output 1 mA Output 2 mALanguage English Display left Display right Out Outputs Output 1 controlOutput Measurement Output 1 Setup Output 1 ControlRange 4-20 mA Dampen 0 sec Hold Last Value Configure Alarm 1 controlAlarm 1 setup Alarm 2 control Changing Alarm ParametersAlarm 1 control Alarm 1 setupAlarm Low Activate ProcessAlarm Fault Feed limit DisableTimeout 3600 sec Alarm 4 setupAlarm 4 setup Feed limit timer Interval timerTimer Disable Timer Time activated Interval 24.0 hr Interval timer setupTemperature Compensation and Temperature Units Temp comp AutoTemp units C Temp comp Manual Temp units C60 Hz Noise ReductionBar meas Auto Bar units mm HgBarometric Pressure Main sensor cal Salinity 0.0 o/ooNoise rejection Main sensor cal Security SecurityAnalyzer Mode Priority Condition DefinitionsAction Definitions Section Calibration Temperature IntroductionAdjust temp + 025.1 C Temperature CalibrationSection Calibration Dissolved Oxygen Sensor Current as a Function Dissolved Oxygen ConcentrationSensor must be in zero solution Zeroing the SensorStabilizing . . . Wait Calibrating the Sensor in AIRSection Calibrate 8.32 ppm Calibrating the Sensor Against a Standard InstrumentZero main sensor Adjust temperature Calibrate main sensorStandardize Calibrating Barometric PressureSection Calibration Current Outputs Trimming the OutputsSection Maintenance Replacement PartsCleaning the membrane Replacing electrolyte solution and membraneSection Power Fuse Spare PartsReplacement Parts Location Part Number Description AIR Blast Sensor WasherSection Troubleshooting Troubleshooting When a Fault Message is ShowingFault message Explanation See Section OverviewTemperature error low or high High input currentCheck sensor zero Sense line openTroubleshooting When no Fault Message is Showing Oxygen Problem See SectionZero current is too high Possible error warning during in-process calibration Zero reading Is unstableBarometric pressure reading is too high or too low Process readings are erratic Readings driftSensor does not respond to changes in oxygen level Oxygen readings are too lowTroubleshooting not Related to Measurement Problems Problem ActionSimulating Inputs Dissolved Oxygen Sensor Polarizing Voltage Resistance Expected currentSimulating Temperature Simulating temperatureSection Return of Material GeneralReturn of Materials Request Sensor or Circuit Board onlyWarranty Specifications subject to change without notice