Emerson 51-DO-03, 51-DO-04 instruction manual Section

Page 52

MODEL DO-03/04

SECTION 7.0

 

CALIBRATION - DISSOLVED OXYGEN

8. Press Exit (F1) four times to return to the main display.

NOTE

If Hold was activated during calibration, “Hold Mode Activated” will continue to flash in the main display. Return the sensor to normal and deactivate Hold. Refer to Section 5.5, step 9.

9.During calibration, the analyzer stores the measured current and calcu- lates the sensitivity. Sensitivity is the sensor current in nA divided by the

saturation concentration of oxygen in ppm. Typical sensitivity for a 499ADO sensor is 1,800 to 3,100 nA/ppm (15 - 25 μA in air).

To view the sensitivity from the main display, press any key to enter the main menu. Press the ￿ key once. Then press Enter (F4) to display the

diagnostic variables. The sensitivity is the third line on the screen. Note the units: nA is nanoamps, μA is microamps.

NOTE

During calibration, ERROR and WARNING messages may appear. If an ERROR message appears, press Exit (F1) to leave and return to the previous screen. If a WARNING message appears, press Cont (F3) to continue the calibration or press Abort (F1) to leave. Continuing the calibration after a warning message appears may cause substantial errors in the subsequent measurement. Refer to Section 15.4 for assis- tance.

46

Image 52
Contents Dissolved Oxygen Measurement System with Air Blast Cleaner Model DO-03/04Essential Instructions Page Model DO-03/04 Measuring System Section Title10.0 List of Tables11.0 Section Title List of FiguresIii Section Specifications Features and BenefitsPower Repeatability ±0.05 ppm 25CRFI/EMI EN-61326 Output Accuracy ± 0.05 mAMaximum distance between compressor and sensor Maximum continuous pressure 50 psig 446 kPa absMaterials of Construction Analyzer dimensions Enclosure dimensions Description Power required selection9240048-00 Tag, stainless steel, specify marking Section Installation InstallationUnpacking and Inspection Model or part number DescriptionInstalling the handrail mounting assembly Installing the air compressor enclosureAssembling the pipe and pipe clamp Attaching the pipe boom to the handrail mounting assembly Installing the sensor in the washer head assemblyAssembling the washer head and pipe boom Attaching the air compressor enclosure to the handrail Installing the sensor in the washer head assembly POWER, ALARM, and Output Wiring Section WiringGeneral 454EPH02 Section Section Wiring the Oxygen Sensor Sensor WiringGeneral Alarm Status Section Display and OperationDisplay KEY Functions and ControlSection Software Configuration Menu Tree for the 54eA AnalyzerOn following Main Menu Calibrate See Diagnostic Variable see Program Settings List Choices Factory Settings SetpointsChoices Factory Settings Configure Choices Factory Settings Alarm setpoints Alarm 1 setpointAlarm 2 setpoint Alarm 3 setpointMA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mA Ranging the OutputsSimulated tests Test alarm 1 OpenTest output Outputs Alarms DisplayMeasure Oxygen Meas units ppmTemp units C Output 1 mA Output 2 mA Display contrast Timeout OnLanguage English Display left Display right Out Output 1 control OutputsOutput Measurement Output 1 Control Output 1 SetupRange 4-20 mA Dampen 0 sec Hold Last Value Configure Alarm 1 controlAlarm 1 setup Alarm 2 control Changing Alarm ParametersAlarm 1 control Alarm 1 setupAlarm Low Activate ProcessAlarm Fault Feed limit DisableTimeout 3600 sec Alarm 4 setupAlarm 4 setup Feed limit timer Interval timerTimer Disable Timer Time activated Interval 24.0 hr Interval timer setupTemperature Compensation and Temperature Units Temp comp AutoTemp units C Temp comp Manual Temp units C60 Hz Noise ReductionBar units mm Hg Bar meas AutoBarometric Pressure Main sensor cal Salinity 0.0 o/ooNoise rejection Main sensor cal Security SecurityCondition Definitions Analyzer Mode PriorityAction Definitions Section Calibration Temperature IntroductionAdjust temp + 025.1 C Temperature CalibrationSection Calibration Dissolved Oxygen Sensor Current as a Function Dissolved Oxygen ConcentrationSensor must be in zero solution Zeroing the SensorStabilizing . . . Wait Calibrating the Sensor in AIRSection Calibrate 8.32 ppm Calibrating the Sensor Against a Standard InstrumentZero main sensor Adjust temperature Calibrate main sensorStandardize Calibrating Barometric PressureSection Calibration Current Outputs Trimming the OutputsSection Maintenance Replacement PartsCleaning the membrane Replacing electrolyte solution and membraneSection Power Fuse Spare PartsReplacement Parts Location Part Number Description AIR Blast Sensor WasherSection Troubleshooting Troubleshooting When a Fault Message is ShowingFault message Explanation See Section OverviewTemperature error low or high High input currentCheck sensor zero Sense line openProblem See Section Troubleshooting When no Fault Message is Showing OxygenZero current is too high Zero reading Is unstable Possible error warning during in-process calibrationBarometric pressure reading is too high or too low Process readings are erratic Readings driftSensor does not respond to changes in oxygen level Oxygen readings are too lowTroubleshooting not Related to Measurement Problems Problem ActionSimulating Inputs Dissolved Oxygen Sensor Polarizing Voltage Resistance Expected currentSimulating Temperature Simulating temperatureSection Return of Material GeneralReturn of Materials Request Sensor or Circuit Board onlyWarranty Specifications subject to change without notice