MODEL | SECTION 7.0 |
| CALIBRATION - DISSOLVED OXYGEN |
8. Press Exit (F1) four times to return to the main display.
NOTE
If Hold was activated during calibration, “Hold Mode Activated” will continue to flash in the main display. Return the sensor to normal and deactivate Hold. Refer to Section 5.5, step 9.
9.During calibration, the analyzer stores the measured current and calcu- lates the sensitivity. Sensitivity is the sensor current in nA divided by the
saturation concentration of oxygen in ppm. Typical sensitivity for a 499ADO sensor is 1,800 to 3,100 nA/ppm (15 - 25 μA in air).
To view the sensitivity from the main display, press any key to enter the main menu. Press the key once. Then press Enter (F4) to display the
diagnostic variables. The sensitivity is the third line on the screen. Note the units: nA is nanoamps, μA is microamps.
NOTE
During calibration, ERROR and WARNING messages may appear. If an ERROR message appears, press Exit (F1) to leave and return to the previous screen. If a WARNING message appears, press Cont (F3) to continue the calibration or press Abort (F1) to leave. Continuing the calibration after a warning message appears may cause substantial errors in the subsequent measurement. Refer to Section 15.4 for assis- tance.
46