MODEL | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
5.4 CHOOSING DISPLAY OPTIONS
1.The 54eA analyzer can be used with other oxygen sensors manufactured by Rosemount Analytical as well as with chlorine, monochloramine, and ozone sensors. The user must configure the analyzer to match the sensor being used.
2.The display menu also lets the user customize the third line in the display, change timeout values, choose a lan- guage other than English, and change the display contrast.
3.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).
Output setpoints Simulated tests Configure
4. Move the cursor to "Configure" and press Enter (F4).
ExitEnter
Display
Outputs
Alarms
5. With the cursor on "Display", press Enter (F4).
ExitEnter
Measure : Oxygen
Sensor: RMT Standard
Meas units: ppm
ExitEdit
6.The screen at left appears. The default settings shown for “Measure” (Oxygen) and “Sensor” (RMT Standard) are appropriate for the 499ADO sensor. There are no changes to make.
If you wish to change the units, use the key to move the cursor to "Meas units" and press Edit (F4). Use the or keys to scroll through the choices. Press Save (F4) to store the selection.
27