Emerson 51-DO-03, 51-DO-04 instruction manual Outputs, Output 1 control, Output Measurement

Page 35

MODEL DO-03/04

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.5 CHANGING OUTPUT PARAMETERS

1.This section describes how to configure the analyzer outputs. Outputs can be configured to represent dis- solved oxygen or temperature.

Output setpoints Simulated tests Configure

2.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).

3.Move the cursor to "Configure" and press Enter (F4).

ExitEnter

Display

Outputs

Alarms

4. Move the cursor to "Outputs" and press Enter (F4).

ExitEnter

Output 1 control

Output 1 setup

Output 2 control

ExitEnter

Output Measurement

Control Mode

5.Five menu headers relate to outputs. Each output has a control head- er and a setup header. The fifth header allows the output hold feature to be configured.

NOTE

The 54eA analyzer can be ordered with optional PID control feature. The version of the 54eA analyzer provided with the DO-03/04 does not have control. Although the control head- er appears in this screen, it is not activated.

To access a header, move the cursor to the desired header and press Enter (F4).

6.Output Control Settings:

a.Move the cursor to the desired output control header. Press Enter (F4).

b.With the cursor on "Output Measurement" press Enter (F4).

ExitEnter

c.Press Edit (F4).

Output : Process

d.Use the ￿ key to scroll through the choices: "Process" and "Temperature". "Process" means dissolved oxygen. Press Save (F4) to store the selection.

Exit

Edit

e. The display returns to the “Output: Process” screen. Press Exit

 

 

 

 

(F1) until the screen in Step 5 appears.

29

Image 35
Contents Model DO-03/04 Dissolved Oxygen Measurement System with Air Blast CleanerEssential Instructions Page Section Title Model DO-03/04 Measuring System11.0 List of Tables10.0 Iii List of FiguresSection Title Features and Benefits Section SpecificationsOutput Accuracy ± 0.05 mA PowerRepeatability ±0.05 ppm 25C RFI/EMI EN-61326Materials of Construction Maximum continuous pressure 50 psig 446 kPa absMaximum distance between compressor and sensor Analyzer dimensions Enclosure dimensions 9240048-00 Tag, stainless steel, specify marking Power required selectionDescription Model or part number Description Section InstallationInstallation Unpacking and InspectionAssembling the pipe and pipe clamp Installing the air compressor enclosureInstalling the handrail mounting assembly Assembling the washer head and pipe boom Installing the sensor in the washer head assemblyAttaching the pipe boom to the handrail mounting assembly Attaching the air compressor enclosure to the handrail Installing the sensor in the washer head assembly General Section WiringPOWER, ALARM, and Output Wiring 454EPH02 Section Section General Sensor WiringWiring the Oxygen Sensor KEY Functions and Control Alarm StatusSection Display and Operation DisplayMenu Tree for the 54eA Analyzer Section Software ConfigurationOn following Main Menu Calibrate See Diagnostic Variable see Choices Factory Settings Setpoints Program Settings ListChoices Factory Settings Configure Choices Factory Settings Alarm 3 setpoint Alarm setpointsAlarm 1 setpoint Alarm 2 setpointRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mATest output Test alarm 1 OpenSimulated tests Meas units ppm Outputs AlarmsDisplay Measure OxygenLanguage English Display left Display right Out Display contrast Timeout OnTemp units C Output 1 mA Output 2 mA Output Measurement OutputsOutput 1 control Range 4-20 mA Dampen 0 sec Hold Last Value Output 1 SetupOutput 1 Control Changing Alarm Parameters ConfigureAlarm 1 control Alarm 1 setup Alarm 2 controlActivate Process Alarm 1 controlAlarm 1 setup Alarm LowAlarm 4 setup Alarm FaultFeed limit Disable Timeout 3600 secInterval timer setup Alarm 4 setup Feed limit timerInterval timer Timer Disable Timer Time activated Interval 24.0 hrTemp comp Manual Temp units C Temperature Compensation and Temperature UnitsTemp comp Auto Temp units CNoise Reduction 60 HzBarometric Pressure Bar meas AutoBar units mm Hg Salinity 0.0 o/oo Main sensor calSecurity Noise rejection Main sensor cal SecurityAction Definitions Analyzer Mode PriorityCondition Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CSensor Current as a Function Dissolved Oxygen Concentration Section Calibration Dissolved OxygenZeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrating Barometric Pressure Zero main sensor Adjust temperatureCalibrate main sensor StandardizeTrimming the Outputs Section Calibration Current OutputsReplacing electrolyte solution and membrane Section MaintenanceReplacement Parts Cleaning the membraneSection Spare Parts Power FuseAIR Blast Sensor Washer Replacement Parts Location Part Number DescriptionOverview Section TroubleshootingTroubleshooting When a Fault Message is Showing Fault message Explanation See SectionSense line open Temperature error low or highHigh input current Check sensor zeroZero current is too high Troubleshooting When no Fault Message is Showing OxygenProblem See Section Barometric pressure reading is too high or too low Possible error warning during in-process calibrationZero reading Is unstable Oxygen readings are too low Process readings are erraticReadings drift Sensor does not respond to changes in oxygen levelSensor Polarizing Voltage Resistance Expected current Troubleshooting not Related to Measurement ProblemsProblem Action Simulating Inputs Dissolved OxygenSimulating temperature Simulating TemperatureGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestWarranty Specifications subject to change without notice