Emerson 51-DO-03, 51-DO-04 Bar meas Auto, Bar units mm Hg, Barometric Pressure

Page 43

MODEL DO-03/04

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.9BAROMETRIC PRESSURE

1.Barometric pressure is used during air calibration.

Noise rejection Barometric pressure Main sensor cal

ExitEnter

Bar meas : Auto

Bar units: mm Hg

ExitEdit

2.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).

3.Move the cursor to "Configure" and press Enter (F4).

4.Move the cursor to "Barometric pressure" and press Enter (F4).

5.Use the ￿ and ￿ keys to move through the list of items. To make a change, press Edit (F4). Use the arrow keys to change settings to the desired value and press Save (F4).

Bar meas: The analyzer has an on-board sensor that automatically measures barometric pressure. To bypass the pressure sensor, select "Manual'. Be sure to enter the desired pressure in the third line. During air calibration, the analyzer will use the barometric pressure entered in this step no matter what the true pressure is.

Bar units: Select the units in which barometric pressure will be dis- played.

37

Image 43
Contents Model DO-03/04 Dissolved Oxygen Measurement System with Air Blast CleanerEssential Instructions Page Section Title Model DO-03/04 Measuring System10.0 List of Tables11.0 Section Title List of FiguresIii Features and Benefits Section SpecificationsOutput Accuracy ± 0.05 mA PowerRepeatability ±0.05 ppm 25C RFI/EMI EN-61326Maximum distance between compressor and sensor Maximum continuous pressure 50 psig 446 kPa absMaterials of Construction Analyzer dimensions Enclosure dimensions Description Power required selection9240048-00 Tag, stainless steel, specify marking Model or part number Description Section InstallationInstallation Unpacking and InspectionInstalling the handrail mounting assembly Installing the air compressor enclosureAssembling the pipe and pipe clamp Attaching the pipe boom to the handrail mounting assembly Installing the sensor in the washer head assemblyAssembling the washer head and pipe boom Attaching the air compressor enclosure to the handrail Installing the sensor in the washer head assembly POWER, ALARM, and Output Wiring Section WiringGeneral 454EPH02 Section Section Wiring the Oxygen Sensor Sensor WiringGeneral KEY Functions and Control Alarm StatusSection Display and Operation DisplayMenu Tree for the 54eA Analyzer Section Software ConfigurationOn following Main Menu Calibrate See Diagnostic Variable see Choices Factory Settings Setpoints Program Settings ListChoices Factory Settings Configure Choices Factory Settings Alarm 3 setpoint Alarm setpointsAlarm 1 setpoint Alarm 2 setpointRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mASimulated tests Test alarm 1 OpenTest output Meas units ppm Outputs AlarmsDisplay Measure OxygenTemp units C Output 1 mA Output 2 mA Display contrast Timeout OnLanguage English Display left Display right Out Output 1 control OutputsOutput Measurement Output 1 Control Output 1 SetupRange 4-20 mA Dampen 0 sec Hold Last Value Changing Alarm Parameters ConfigureAlarm 1 control Alarm 1 setup Alarm 2 controlActivate Process Alarm 1 controlAlarm 1 setup Alarm LowAlarm 4 setup Alarm FaultFeed limit Disable Timeout 3600 secInterval timer setup Alarm 4 setup Feed limit timerInterval timer Timer Disable Timer Time activated Interval 24.0 hrTemp comp Manual Temp units C Temperature Compensation and Temperature UnitsTemp comp Auto Temp units CNoise Reduction 60 HzBar units mm Hg Bar meas AutoBarometric Pressure Salinity 0.0 o/oo Main sensor calSecurity Noise rejection Main sensor cal SecurityCondition Definitions Analyzer Mode PriorityAction Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CSensor Current as a Function Dissolved Oxygen Concentration Section Calibration Dissolved OxygenZeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrating Barometric Pressure Zero main sensor Adjust temperatureCalibrate main sensor StandardizeTrimming the Outputs Section Calibration Current OutputsReplacing electrolyte solution and membrane Section MaintenanceReplacement Parts Cleaning the membraneSection Spare Parts Power FuseAIR Blast Sensor Washer Replacement Parts Location Part Number DescriptionOverview Section TroubleshootingTroubleshooting When a Fault Message is Showing Fault message Explanation See SectionSense line open Temperature error low or highHigh input current Check sensor zeroProblem See Section Troubleshooting When no Fault Message is Showing OxygenZero current is too high Zero reading Is unstable Possible error warning during in-process calibrationBarometric pressure reading is too high or too low Oxygen readings are too low Process readings are erraticReadings drift Sensor does not respond to changes in oxygen levelSensor Polarizing Voltage Resistance Expected current Troubleshooting not Related to Measurement ProblemsProblem Action Simulating Inputs Dissolved OxygenSimulating temperature Simulating TemperatureGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestWarranty Specifications subject to change without notice