Emerson 51-DO-03, 51-DO-04 instruction manual Section Calibration Dissolved Oxygen

Page 49

MODEL DO-03/04

SECTION 7.0

 

CALIBRATION - DISSOLVED OXYGEN

SECTION 7.0

CALIBRATION - DISSOLVED OXYGEN

7.1 INTRODUCTION

As Figure 7-1 shows, oxygen sensors generate a current directly proportional to the concentration of dissolved oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero stan- dard) and to a solution containing a known amount of oxygen (full-scale standard).

The zero standard is necessary because oxygen sensors, even when no oxygen is present in the sample, gener- ate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a dissolved oxygen value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The recommended zero standard is 5% sodium sulfite in water, although oxygen-free nitrogen can also be used.

The purpose of the full-scale standard is to establish the slope of the calibration curve. Because the solubility of atmospheric oxygen in water as a function of temperature and barometric pressure is well known, the natural choice for a full-scale standard is air-saturated water. However, air-saturated water is difficult to prepare and use, so the universal practice is to use air for calibration. From the point of view of the oxygen sensor, air and air-sat- urated water are identical. The equivalence comes about because the sensor really measures the chemical poten- tial of oxygen. Chemical potential is the force that causes oxygen molecules to diffuse from the sample into the sensor where they can be measured. It is also the force that causes oxygen molecules in air to dissolve in water and to continue to dissolve until the water is saturated with oxygen. Once the water is saturated, the chemical potential of oxygen in the two phases (air and water) is the same.

Oxygen sensors generate a current directly proportional to the rate at which oxygen molecules diffuse through a membrane stretched over the end of the sensor. The diffusion rate depends on the difference in chemical poten- tial between oxygen in the sensor and oxygen in the sample. An electrochemical reaction, which destroys any oxy- gen molecules entering the sensor, keeps the concentration (and the chemical potential) of oxygen inside the sen- sor equal to zero. Therefore, the chemical potential of oxygen in the sample alone determines the diffusion rate and the sensor current.

When the sensor is calibrated, the chemical potential of oxygen in the standard determines the sensor current. Whether the sensor is calibrated in air or air-saturated water is immaterial. The chemical potential of oxygen is the same in either phase. Normally, to make the calculation of solubility in common units (like ppm DO) simpler, it is convenient to use water-saturated air for calibration.

Automatic air calibration is standard. The user simply exposes the sensor to water-saturated air. The analyzer monitors the sensor current. When the current is stable, the analyzer stores the current and measures the baro- metric pressure and temperature. The temperature element is part of the dissolved oxygen sensor. The pressure sensor is inside the analyzer. From the temperature, the analyzer calculates the saturation vapor pressure of water. Next, it calculates the pressure of dry air by subtracting the vapor pressure from the barometric pressure. Using the fact that dry air always contains 20.95% oxygen, the

analyzer calculates the partial pressure of oxygen. Once the analyzer knows the partial pressure of oxygen, it uses the Bunsen coefficient to calculate the equilibrium solubil- ity of atmospheric oxygen in water at the prevailing tem- perature. At 25°C and 760 mm Hg, the equilibrium solu- bility is 8.24 ppm.

Often it is too difficult or messy to remove the sensor from the process liquid for calibration. In this case, the sensor can be calibrated against a measurement made with a portable laboratory instrument. The laboratory instrument typically uses a membrane-covered amperometric sensor that has been calibrated against water-saturated air.

FIGURE 7-1. Sensor Current as a Function of

Dissolved Oxygen Concentration

43

Image 49
Contents Model DO-03/04 Dissolved Oxygen Measurement System with Air Blast CleanerEssential Instructions Page Section Title Model DO-03/04 Measuring System10.0 List of Tables11.0 Section Title List of FiguresIii Features and Benefits Section SpecificationsRepeatability ±0.05 ppm 25C PowerRFI/EMI EN-61326 Output Accuracy ± 0.05 mAMaximum distance between compressor and sensor Maximum continuous pressure 50 psig 446 kPa absMaterials of Construction Analyzer dimensions Enclosure dimensions Description Power required selection9240048-00 Tag, stainless steel, specify marking Installation Section InstallationUnpacking and Inspection Model or part number DescriptionInstalling the handrail mounting assembly Installing the air compressor enclosureAssembling the pipe and pipe clamp Attaching the pipe boom to the handrail mounting assembly Installing the sensor in the washer head assemblyAssembling the washer head and pipe boom Attaching the air compressor enclosure to the handrail Installing the sensor in the washer head assembly POWER, ALARM, and Output Wiring Section WiringGeneral 454EPH02 Section Section Wiring the Oxygen Sensor Sensor WiringGeneral Section Display and Operation Alarm StatusDisplay KEY Functions and ControlMenu Tree for the 54eA Analyzer Section Software ConfigurationOn following Main Menu Calibrate See Diagnostic Variable see Choices Factory Settings Setpoints Program Settings ListChoices Factory Settings Configure Choices Factory Settings Alarm 1 setpoint Alarm setpointsAlarm 2 setpoint Alarm 3 setpointRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mASimulated tests Test alarm 1 OpenTest output Display Outputs AlarmsMeasure Oxygen Meas units ppmTemp units C Output 1 mA Output 2 mA Display contrast Timeout OnLanguage English Display left Display right Out Output 1 control OutputsOutput Measurement Output 1 Control Output 1 SetupRange 4-20 mA Dampen 0 sec Hold Last Value Alarm 1 control ConfigureAlarm 1 setup Alarm 2 control Changing Alarm ParametersAlarm 1 setup Alarm 1 controlAlarm Low Activate ProcessFeed limit Disable Alarm FaultTimeout 3600 sec Alarm 4 setupInterval timer Alarm 4 setup Feed limit timerTimer Disable Timer Time activated Interval 24.0 hr Interval timer setupTemp comp Auto Temperature Compensation and Temperature UnitsTemp units C Temp comp Manual Temp units CNoise Reduction 60 HzBar units mm Hg Bar meas AutoBarometric Pressure Salinity 0.0 o/oo Main sensor calSecurity Noise rejection Main sensor cal SecurityCondition Definitions Analyzer Mode PriorityAction Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CSensor Current as a Function Dissolved Oxygen Concentration Section Calibration Dissolved OxygenZeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrate main sensor Zero main sensor Adjust temperatureStandardize Calibrating Barometric PressureTrimming the Outputs Section Calibration Current OutputsReplacement Parts Section MaintenanceCleaning the membrane Replacing electrolyte solution and membraneSection Spare Parts Power FuseAIR Blast Sensor Washer Replacement Parts Location Part Number DescriptionTroubleshooting When a Fault Message is Showing Section TroubleshootingFault message Explanation See Section OverviewHigh input current Temperature error low or highCheck sensor zero Sense line openProblem See Section Troubleshooting When no Fault Message is Showing OxygenZero current is too high Zero reading Is unstable Possible error warning during in-process calibrationBarometric pressure reading is too high or too low Readings drift Process readings are erraticSensor does not respond to changes in oxygen level Oxygen readings are too lowProblem Action Troubleshooting not Related to Measurement ProblemsSimulating Inputs Dissolved Oxygen Sensor Polarizing Voltage Resistance Expected currentSimulating temperature Simulating TemperatureGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestWarranty Specifications subject to change without notice