Emerson DO-04, 51-DO-03 Calibrate 8.32 ppm, Calibrating the Sensor Against a Standard Instrument

Page 53

MODEL DO-03/04

SECTION 7.0

 

CALIBRATION - DISSOLVED OXYGEN

7.4 CALIBRATING THE SENSOR AGAINST A STANDARD INSTRUMENT

The analyzer and sensor can be calibrated against a standard instrument. For oxygen sensors installed in aera- tion basins in waste treatment plants, calibration against a second instrument is often preferred. For an accurate calibration be sure that…

1.The standard instrument has been zeroed and calibrated against water-saturated air following the manufac- turer's instructions.

2.The standard sensor is immersed in the liquid as close to the process sensor as possible.

3.Adequate time is allowed for the standard sensor to stabilize before calibrating the process instrument.

Calibrate main sensor Zero main sensor Adjust temperature

4.From the main display, press any key to obtain the main menu. With the cursor on “Calibrate,” press Enter (F4).

NOTE

If Hold was enabled in Section 5.5, the hold screen will appear. To activate hold, refer to Section 5.5, step 9.

ExitEnter

Calibrate in air Calibrate in process Barometric pressure

ExitEnter

8.32ppm

Stabilizing . . . Wait

Abort Cont

8.32ppm

Calibrate : 8.32 ppm

ExitEdit

Press Enter (F4).

5.The screen at left appears. Press the ￿ key once to move the cursor to “Calibrate in process.” Then, press Enter (F4).

6.Press Cont (F3). “Wait” flashes until the sensor is stabilized.

If the analyzer appears locked, the reading is not stable enough. Increase the amount the concentration is permitted to vary or decrease the stabilization time. See Section 5.10. Repeat the calibra- tion. To proceed with the calibration as is, press Cont (F3).

7.Once the reading is stable, the screen at left appears. Press Edit (F4). Use the arrow keys to change the concentration in the second line of the display to the desired value. Press Save (F4) to store the value.

8.Press Exit (F1) four times to return to the main display.

NOTE

If Hold was activated during calibration, “Hold Mode Activated” will continue to flash in the main display. Return the sensor to normal and deactivate Hold. Refer to Section 5.5, step 9.

47

Image 53
Contents Model DO-03/04 Dissolved Oxygen Measurement System with Air Blast CleanerEssential Instructions Page Section Title Model DO-03/04 Measuring System11.0 List of Tables10.0 Iii List of FiguresSection Title Features and Benefits Section SpecificationsRepeatability ±0.05 ppm 25C PowerRFI/EMI EN-61326 Output Accuracy ± 0.05 mAMaterials of Construction Maximum continuous pressure 50 psig 446 kPa absMaximum distance between compressor and sensor Analyzer dimensions Enclosure dimensions 9240048-00 Tag, stainless steel, specify marking Power required selectionDescription Installation Section InstallationUnpacking and Inspection Model or part number DescriptionAssembling the pipe and pipe clamp Installing the air compressor enclosureInstalling the handrail mounting assembly Assembling the washer head and pipe boom Installing the sensor in the washer head assemblyAttaching the pipe boom to the handrail mounting assembly Attaching the air compressor enclosure to the handrail Installing the sensor in the washer head assembly General Section WiringPOWER, ALARM, and Output Wiring 454EPH02 Section Section General Sensor WiringWiring the Oxygen Sensor Section Display and Operation Alarm StatusDisplay KEY Functions and ControlMenu Tree for the 54eA Analyzer Section Software ConfigurationOn following Main Menu Calibrate See Diagnostic Variable see Choices Factory Settings Setpoints Program Settings ListChoices Factory Settings Configure Choices Factory Settings Alarm 1 setpoint Alarm setpointsAlarm 2 setpoint Alarm 3 setpointRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mATest output Test alarm 1 OpenSimulated tests Display Outputs AlarmsMeasure Oxygen Meas units ppmLanguage English Display left Display right Out Display contrast Timeout OnTemp units C Output 1 mA Output 2 mA Output Measurement OutputsOutput 1 control Range 4-20 mA Dampen 0 sec Hold Last Value Output 1 SetupOutput 1 Control Alarm 1 control ConfigureAlarm 1 setup Alarm 2 control Changing Alarm ParametersAlarm 1 setup Alarm 1 controlAlarm Low Activate ProcessFeed limit Disable Alarm FaultTimeout 3600 sec Alarm 4 setupInterval timer Alarm 4 setup Feed limit timerTimer Disable Timer Time activated Interval 24.0 hr Interval timer setupTemp comp Auto Temperature Compensation and Temperature UnitsTemp units C Temp comp Manual Temp units CNoise Reduction 60 HzBarometric Pressure Bar meas AutoBar units mm Hg Salinity 0.0 o/oo Main sensor calSecurity Noise rejection Main sensor cal SecurityAction Definitions Analyzer Mode PriorityCondition Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CSensor Current as a Function Dissolved Oxygen Concentration Section Calibration Dissolved Oxygen Zeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrate main sensor Zero main sensor Adjust temperatureStandardize Calibrating Barometric PressureTrimming the Outputs Section Calibration Current OutputsReplacement Parts Section MaintenanceCleaning the membrane Replacing electrolyte solution and membraneSection Spare Parts Power FuseAIR Blast Sensor Washer Replacement Parts Location Part Number DescriptionTroubleshooting When a Fault Message is Showing Section TroubleshootingFault message Explanation See Section OverviewHigh input current Temperature error low or highCheck sensor zero Sense line openZero current is too high Troubleshooting When no Fault Message is Showing OxygenProblem See Section Barometric pressure reading is too high or too low Possible error warning during in-process calibrationZero reading Is unstable Readings drift Process readings are erraticSensor does not respond to changes in oxygen level Oxygen readings are too lowProblem Action Troubleshooting not Related to Measurement ProblemsSimulating Inputs Dissolved Oxygen Sensor Polarizing Voltage Resistance Expected currentSimulating temperature Simulating TemperatureGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestWarranty Specifications subject to change without notice