Emerson DO-04, 51-DO-03 Troubleshooting not Related to Measurement Problems, Problem Action

Page 65

MODEL DO-03/04

 

SECTION 10.0

 

 

 

 

TROUBLESHOOTING

 

10.5 TROUBLESHOOTING NOT RELATED TO MEASUREMENT PROBLEMS

 

 

 

 

 

 

Problem

Action

 

 

Display segments missing

Replace display board

 

 

Alarm relays are chattering

1.

Check alarm setpoints.

 

 

 

2. Increase hysteresis time delay settings (see Section 5.6)

 

 

Incorrect current output

1.

Verify that output load is less than 600 Ω.

 

 

 

2.

For minor errors, trim outputs (see Section 8.0)

 

 

 

3.

Replace power supply board

 

 

Display too light or too dark

Change contrast (see Section 5.4)

 

 

“Level 1, 2 or 3 security: Lock” shown in display

Controller has password protection (see Section 5.11)

 

 

“Hold mode activated” showing in display

Controller is in hold (see Section 5.5, steps 8 and 9)

 

 

“Simulating output 1 or 2” showing in display

Controller is simulating outputs (see Section 5.3)

 

 

“Simulating alarm 1, 2, 3 or 4” showing in display

Controller is simulating alarms (see Section 5.3)

 

 

 

 

 

 

10.6 SIMULATING INPUTS - DISSOLVED OXYGEN

To check the performance of the controller, use a decade box to simulate the current from the oxygen sensor.

A. Disconnect the anode and cathode leads from terminals 1 and 2 on TB3 and connect a decade box as shown in Figure 10-1. It is not necessary to disconnect the RTD leads.

B. Set the decade box to the resistance shown in the table.

 

 

Sensor

 

Polarizing Voltage

 

Resistance

Expected current

 

 

 

 

499ADO

 

-675 mV

 

34 kΩ

20 μA

 

 

 

 

 

 

 

 

 

 

 

C. Note the sensor current. To view the sensor current from the

 

 

 

 

 

 

 

 

main display, press any key to enter the main menu. Move the

 

 

 

cursor to "Diagnostics" and press Enter (F4). The sensor cur-

 

 

 

rent is the second line in the display. Note the units: μA is

 

 

 

microamps, nA is nanoamps.

 

 

 

 

 

D. Change the decade box resistance and verify that the correct

 

 

 

current is shown. Calculate current from the equation:

 

 

 

 

 

 

 

current (μA) =

voltage (mV)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

resistance (kΩ)

 

 

 

 

 

FIGURE 10-1. Simulate Dissolved

Oxygen

59

Image 65
Contents Model DO-03/04 Dissolved Oxygen Measurement System with Air Blast CleanerEssential Instructions Page Section Title Model DO-03/04 Measuring System11.0 List of Tables10.0 Iii List of FiguresSection Title Features and Benefits Section SpecificationsRepeatability ±0.05 ppm 25C PowerRFI/EMI EN-61326 Output Accuracy ± 0.05 mAMaterials of Construction Maximum continuous pressure 50 psig 446 kPa absMaximum distance between compressor and sensor Analyzer dimensions Enclosure dimensions 9240048-00 Tag, stainless steel, specify marking Power required selectionDescription Installation Section InstallationUnpacking and Inspection Model or part number DescriptionAssembling the pipe and pipe clamp Installing the air compressor enclosureInstalling the handrail mounting assembly Assembling the washer head and pipe boom Installing the sensor in the washer head assemblyAttaching the pipe boom to the handrail mounting assembly Attaching the air compressor enclosure to the handrail Installing the sensor in the washer head assembly General Section WiringPOWER, ALARM, and Output Wiring 454EPH02 Section Section General Sensor WiringWiring the Oxygen Sensor Section Display and Operation Alarm StatusDisplay KEY Functions and ControlMenu Tree for the 54eA Analyzer Section Software ConfigurationOn following Main Menu Calibrate See Diagnostic Variable see Choices Factory Settings Setpoints Program Settings ListChoices Factory Settings Configure Choices Factory Settings Alarm 1 setpoint Alarm setpointsAlarm 2 setpoint Alarm 3 setpointRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mATest output Test alarm 1 OpenSimulated tests Display Outputs AlarmsMeasure Oxygen Meas units ppmLanguage English Display left Display right Out Display contrast Timeout OnTemp units C Output 1 mA Output 2 mA Output Measurement OutputsOutput 1 control Range 4-20 mA Dampen 0 sec Hold Last Value Output 1 SetupOutput 1 Control Alarm 1 control ConfigureAlarm 1 setup Alarm 2 control Changing Alarm ParametersAlarm 1 setup Alarm 1 controlAlarm Low Activate ProcessFeed limit Disable Alarm FaultTimeout 3600 sec Alarm 4 setupInterval timer Alarm 4 setup Feed limit timerTimer Disable Timer Time activated Interval 24.0 hr Interval timer setupTemp comp Auto Temperature Compensation and Temperature UnitsTemp units C Temp comp Manual Temp units CNoise Reduction 60 HzBarometric Pressure Bar meas AutoBar units mm Hg Salinity 0.0 o/oo Main sensor calSecurity Noise rejection Main sensor cal SecurityAction Definitions Analyzer Mode PriorityCondition Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CSensor Current as a Function Dissolved Oxygen Concentration Section Calibration Dissolved OxygenZeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrate main sensor Zero main sensor Adjust temperatureStandardize Calibrating Barometric PressureTrimming the Outputs Section Calibration Current OutputsReplacement Parts Section MaintenanceCleaning the membrane Replacing electrolyte solution and membraneSection Spare Parts Power FuseAIR Blast Sensor Washer Replacement Parts Location Part Number DescriptionTroubleshooting When a Fault Message is Showing Section TroubleshootingFault message Explanation See Section OverviewHigh input current Temperature error low or highCheck sensor zero Sense line openZero current is too high Troubleshooting When no Fault Message is Showing OxygenProblem See Section Barometric pressure reading is too high or too low Possible error warning during in-process calibrationZero reading Is unstable Readings drift Process readings are erraticSensor does not respond to changes in oxygen level Oxygen readings are too lowProblem Action Troubleshooting not Related to Measurement ProblemsSimulating Inputs Dissolved Oxygen Sensor Polarizing Voltage Resistance Expected currentSimulating temperature Simulating TemperatureGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestWarranty Specifications subject to change without notice