HP 16532A - Performance Tests
3-6. DC CAL OUTPUT Test
The DC CAL OUTPUT port on the rear panel is used for instrument calibration and probe calibration. Though calibrator accuracy is not specified in the performance specifications, it must be within limits in or- der to provide accurate calibration.
Test Limits:
5.000 V ± 10 mV
Equipment Required:
The following equipment is required for the test.
|
|
| Recommended |
Equipment Required |
| Critical Specification | Model/Part Number |
|
|
| |
Digital Multimeter | 0.1 mV resolution, better than 0.005% accuracy | HP 3458A | |
Cable | 50 | Ω BNC (m to m) | HP 10503A |
Adapter | 50 | Ω BNC (f) to Dual Banana Plug | HP |
|
|
|
|
|
|
|
|
Procedure:
1.Use the
2.On the oscilloscope module set the following parameters in the order given:
| Menu |
| Selection | Setting |
|
|
|
|
|
Calibration |
| Mode |
| Service Calibration |
|
| Procedure |
| DC Cal BNC |
|
| DC Volts |
| 0 V |
|
|
|
|
|
|
|
|
|
|
3.The DVM should read close to 0.0000 V. Record the reading to four decimal places.
V1 = __________ .
4.In the Calibration menu set the DC Volts to 5 V.
5.The DVM should read close to 5.0000 V. Record the reading to four decimal places.
V2 = __________ .
6.In the Calibration menu set the DC Volts to 0 V.
7.Subtract V1 from V2. The difference should be between 4.990 and 5.010 V. Record the reading in the performance test record.