3-11. BANDWIDTH

This test checks the bandwidth of the oscilloscope module.

Specification:

Bandwidth (dc coupled) dc to 250 MHz

Equipment Required:

The following equipment is required for this test. Procedures are based on the model or part number rec- ommended.

 

 

 

Recommended

Equipment Required

 

Critical Specification

Model/Part Number

 

 

 

 

Signal Generator

1

- 250 MHz at approx 170 mVrms

HP 8656B

Power Meter/Sensor

1

- 250 MHz ± 3% accuracy

HP 436/8482A

Power Splitter

Outputs differ by < 0.15 dB

HP 11667B

Cable

Type N (m to m) 24-inch

HP 11500B

Adapter

50 Ω Type N (m) to BNC (m)

HP 1250-0082

 

 

 

 

 

 

 

 

Procedure:

1.With the N cable, connect the signal generator to the power splitter input. Connect the power sensor to one output of the power splitter.

2.Using an N-to-BNC adapter, connect the other power splitter output to the Channel 1 input of the oscilloscope module.

3.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters.

 

Menu

 

Selection

 

Setting

 

 

 

 

 

 

Channel

 

Input

 

1

 

 

 

Probe

 

1:1

 

 

v/Div

 

80 mV

 

 

Offset

 

0 V

 

 

Coupling

 

50

Ω / DC

 

 

s/Div

 

200 ns

Trigger

 

Mode

 

Edge

 

 

Level

 

0 V

Display

 

Mode

 

Average

 

 

Average #

 

32

 

 

 

Grid

 

On

 

 

 

Markers

 

Off

 

 

 

 

 

 

 

 

 

 

 

 

 

3-11

Image 37
Contents HP 16532A Safety Symbols Safety ConsiderationsContents Calibration ServiceSection Replaceable Parts List of Tables Page Table of Contents Page Introduction Safety RequirementsModule Covered by This Manual Accessories Available Accessories SuppliedSpecifications Operating CharacteristicsRise Time2 1.4 ns HP 16532A SpecificationsWaveform Record Length 8000 points Vertical dc Gain Accuracy4 ± 1.5% of full scaleVertical at BNC HP 16532A Operating CharacteristicsTimebase Digitizer TriggeringTrigger Level Range Trigger ModesDisplay Modes Waveform DisplayMeasurement Aids Operating Environment Setup AidsEquipment Critical Specifications Recommended Model USE Page Table of Contents Page Initial Inspection Safety Requirements Power RequirementsModule Installation Preparation for UseProcedure Installation ConsiderationsEndplate Overlap Bottom Sequence Operating Environment Tagging for ServiceStorage PackagingTable of Contents Page Test Record Performance Test ProceduresPerformance Test Interval DC CAL Output Test AC CAL Output Test Equipment Required Critical Specification Model/Part Number SpecificationInput Resistance Page Voltage Measurement Accuracy Voltage Measurement Accuracy Test Div Offset Supply Upper Limit Lower Limit Offset Accuracy Procedure Zero-Input OffsetDiv Offset Supply Min Max DC Input OffsetBandwidth V1MHz Time Measurement Accuracy Menu Selection Setting Trigger Sensitivity RUN Repetitive Performance Test Record Paragraph No Test Results Yes No Page Table of Contents Page Calibration Interval Calibration Protection SwitchMaintaining System Integrity Calibration Instrument WarmupLocation of the PROTECT/UNPROTECT Switch Setting the PROTECT/UNPROTECT Switch to UnprotectEquipment Required for Calibration Loading the Default Calibration FactorsSelf Cal Menu Calibrations Calibrate Delay of the Self Cal menu as followsCalibrate Vert Cal of the Self Cal menu as follows Calibrate the Logic Trigger of the Self Cal menu as follows Calibrate the Time Null of the Self Cal menu as followsExt Trig Null Calibration Connections Calibrate the Ext Trig Null of the Self Cal menu as followsSealing the Mainframe Setting the PROTECT/UNPROTECT Switch to ProtectTable of Contents Page Abbreviations Replaceable Parts ListOrdering Information Exchange AssembliesDirect Mail Order System Reference Designator and Abbreviations Parts Identification HP Part Qty Number Reference DesignatorDescription Mfr Code Mfr Part NumberPage Table of Contents Page Module Block Diagram and The- ory of Operation Introduction Safety RequirementsMain Assembly Theory of Operation Attenuator Theory of OperationHP 16532A Oscilloscope Block Diagram Page Self-Tests Self-Test Access ProcedureLoad Test System Screen Touch the Functional Tests field Functional Tests ScreenA/D Test Run Screen Exit Test System Screen Data Memory Test Self-Test DescriptionsTimebase Test TestTroubleshooting Aids TroubleshootingVerifying the System Clock Page Troubleshooting Sheet Troubleshooting E E T1 6 5 3 2 a Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Oscilloscope Module Removal Procedure Oscilloscope Module Removal and Replacement12. Endplate Overlap Bottom Sequence Oscilloscope Module Replacement ProcedureTrigger Cable Removal Procedure Trigger Cable Removal and ReplacementTrigger Cable Replacement Procedure Attenuator Removal and ReplacementAttenuator Removal Procedure Attenuator Replacement Procedure Certification WarrantyAssistance