HP 16532A - Performance Tests
3-10. OFFSET ACCURACY
This test verifies the offset accuracy.
Specification:
± (1.0% of channel offset + 2.0% of full scale)
Equipment Required:
The following equipment is required for the test. Procedures are based on the model or part numbers rec- ommended.
|
|
| Recommended |
Equipment Required |
| Critical Specification | Model/Part Number |
|
|
| |
DC Power Supply | − 35.000 to + 35.000 Vdc, ± 1 mV resolution | HP 6114A | |
Digital Multimeter | Better than 0.1% accuracy | HP 3458A | |
Cable | 50 | Ω BNC (m to m) | HP 10503A |
Adapter | 50 | Ω BNC (f) to Dual Banana Plug | HP |
|
|
|
|
|
|
|
|
Procedure:
Zero-Input Offset
1.Disconnect all cables and probes from the oscilloscope inputs. In the Waveform Selection menu de- lete channel 2. If channel 1 is not inserted, insert channel 1. Then set up the oscilloscope module according to the following parameters in the order given:
| Menu |
| Selection | Setting |
|
|
|
|
|
Channel |
| Input |
| 1 |
|
| Probe |
| 1:1 |
|
| v/Div |
| 4 V |
|
| Offset |
| 0 V |
|
| Coupling |
| 1 MΩ / DC |
|
| s/Div |
| 500 ns |
Trigger |
| Mode |
| Immediate |
Display |
| Mode |
| Average |
|
| Average # |
| 32 |
|
| Grid |
| On |
|
| Markers |
| On |
|
|
|
|
|
|
|
|
|
|
2.With the DC power supply disconnected from the channel input, make sure the markers are near the center of the display.