HP 1GSa/s Digitizing Oscilloscope 16532A manual Load Test System Screen

Page 71

HP 16532A - Service

3.Install the Composite PV Test software disk in the inactive disk drive.

4.Touch the box labeled Touch box to Load Test System. See figure 6-3.

Note

When the self-test software is loaded, the user operating system is overwritten in the mainframe system RAM.

Figure 6-3. Load Test System Screen

5.From the test system screen in figure 6-4, touch Test System, then touch 1 GSa/s SCOPE D. (For the example shown in figure 6-4, the oscilloscope module is in slot D. If multiple oscilloscope mod- ules are present, select the one to be tested.)

Figure 6-4. Test System Screen

6-5

Image 71
Contents HP 16532A Safety Symbols Safety ConsiderationsContents Section Replaceable Parts ServiceCalibration List of Tables Page Table of Contents Page Module Covered by This Manual Safety RequirementsIntroduction Operating Characteristics Accessories SuppliedAccessories Available SpecificationsVertical dc Gain Accuracy4 ± 1.5% of full scale HP 16532A SpecificationsRise Time2 1.4 ns Waveform Record Length 8000 pointsTimebase HP 16532A Operating CharacteristicsVertical at BNC Trigger Modes TriggeringDigitizer Trigger Level RangeDisplay Modes Waveform DisplayMeasurement Aids Operating Environment Setup AidsEquipment Critical Specifications Recommended Model USE Page Table of Contents Page Preparation for Use Power RequirementsInitial Inspection Safety Requirements Module InstallationProcedure Installation ConsiderationsEndplate Overlap Bottom Sequence Packaging Tagging for ServiceOperating Environment StorageTable of Contents Page Performance Test Interval Performance Test ProceduresTest Record DC CAL Output Test AC CAL Output Test Input Resistance SpecificationEquipment Required Critical Specification Model/Part Number Page Voltage Measurement Accuracy Voltage Measurement Accuracy Test Div Offset Supply Upper Limit Lower Limit Offset Accuracy Procedure Zero-Input OffsetDiv Offset Supply Min Max DC Input OffsetBandwidth V1MHz Time Measurement Accuracy Menu Selection Setting Trigger Sensitivity RUN Repetitive Performance Test Record Paragraph No Test Results Yes No Page Table of Contents Page Maintaining System Integrity Calibration Protection SwitchCalibration Interval Calibration Instrument WarmupLocation of the PROTECT/UNPROTECT Switch Setting the PROTECT/UNPROTECT Switch to UnprotectEquipment Required for Calibration Loading the Default Calibration FactorsCalibrate Vert Cal of the Self Cal menu as follows Calibrate Delay of the Self Cal menu as followsSelf Cal Menu Calibrations Calibrate the Logic Trigger of the Self Cal menu as follows Calibrate the Time Null of the Self Cal menu as followsExt Trig Null Calibration Connections Calibrate the Ext Trig Null of the Self Cal menu as followsSealing the Mainframe Setting the PROTECT/UNPROTECT Switch to ProtectTable of Contents Page Exchange Assemblies Replaceable Parts ListAbbreviations Ordering InformationDirect Mail Order System Reference Designator and Abbreviations Parts Identification Mfr Code Mfr Part Number Reference DesignatorHP Part Qty Number DescriptionPage Table of Contents Page Attenuator Theory of Operation Introduction Safety RequirementsModule Block Diagram and The- ory of Operation Main Assembly Theory of OperationHP 16532A Oscilloscope Block Diagram Page Self-Tests Self-Test Access ProcedureLoad Test System Screen Touch the Functional Tests field Functional Tests ScreenA/D Test Run Screen Exit Test System Screen Test Self-Test DescriptionsData Memory Test Timebase TestTroubleshooting Aids TroubleshootingVerifying the System Clock Page 1 6 5 3 2 a Troubleshooting E E TTroubleshooting Sheet Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Oscilloscope Module Removal Procedure Oscilloscope Module Removal and Replacement12. Endplate Overlap Bottom Sequence Oscilloscope Module Replacement ProcedureTrigger Cable Removal Procedure Trigger Cable Removal and ReplacementAttenuator Removal Procedure Attenuator Removal and ReplacementTrigger Cable Replacement Procedure Attenuator Replacement Procedure Assistance WarrantyCertification