HP 1GSa/s Digitizing Oscilloscope 16532A manual DC Input Offset, Div Offset Supply Min Max

Page 36

HP 16532A - Performance Tests

3.Select RUN - REPETITIVE. Wait approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V ± 320.0 mV as read from the markers voltage field. Record the reading in the performance test record.

4.In the Channel menu, change the vertical sensitivity to 1 V/Div. Select RUN - REPETITIVE and wait

approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V

± 80.0 mV. Record the reading in the performance test record.

5.In the Channel menu, change the vertical sensitivity to 100 mV/Div. Select RUN - REPETITIVE and

wait approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V ± 8.0 mV. Record the reading in the performance test record.

6.In the Channel menu, change the vertical sensitivity to 10 mV/Div. Select RUN - REPETITIVE and

wait approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V ± 800.0 μV. Record the reading in the performance test record.

DC Input Offset

Set the Channel Coupling field to 1MΩ / DC or damage to the equipment will result.

7.Use the BNC-to-banana adapter to connect the BNC cable between the power supply and channel

1input. Monitor the power supply using a voltmeter.

8.Use the following table for the next steps:

V/Div

Offset

Supply

Min

Max

 

 

 

 

 

1 V/Div

35.00 V

35.00 V

35.4 V

34.6 V

100 mV/Div

10.00 V

10.00 V

10.1 V

9.90 V

20 mV/Div

2.00 V

2.00 V

2.02 V

1.98 V

20 mV/Div

+ 2.00 V

+ 2.00 V

+ 1.98 V

+ 2.02 V

100 mV/Div

+ 10.00 V

+ 10.00 V

+ 9.90 V

+ 10.1 V

1 V/Div

+ 35.00 V

+ 35.00 V

+ 34.6 V

+ 35.4 V

 

 

 

 

 

 

 

 

 

 

9.In the Channel menu set the V/Div range and offset per the first line of the table. Set the power supply per the first line of the table.

10.On the oscilloscope, select RUN - REPETITIVE and wait approximately 15 seconds (averaging complete), then STOP.

11.Read the voltage from the markers voltage field. It should be within the limits given in the table. Re- cord the reading in the performance test record.

12.Repeat steps 7 through 10 for the remaining lines of the table using the V/Div range, offset, and sup- ply voltages given in the table.

13.In the Channel menu select channel 2. In the Waveform Selection menu delete channel 1 and insert channel 2.

14.Repeat steps 1 through 12 for channel 2, replacing channel 1 with channel 2 where applicable.

3-10

Image 36
Contents HP 16532A Safety Considerations Safety SymbolsContents Service CalibrationSection Replaceable Parts List of Tables Page Table of Contents Page Safety Requirements IntroductionModule Covered by This Manual Accessories Supplied Accessories AvailableSpecifications Operating CharacteristicsHP 16532A Specifications Rise Time2 1.4 nsWaveform Record Length 8000 points Vertical dc Gain Accuracy4 ± 1.5% of full scaleHP 16532A Operating Characteristics Vertical at BNCTimebase Triggering DigitizerTrigger Level Range Trigger ModesWaveform Display Display ModesMeasurement Aids Setup Aids Operating EnvironmentEquipment Critical Specifications Recommended Model USE Page Table of Contents Page Power Requirements Initial Inspection Safety RequirementsModule Installation Preparation for UseInstallation Considerations ProcedureEndplate Overlap Bottom Sequence Tagging for Service Operating EnvironmentStorage PackagingTable of Contents Page Performance Test Procedures Test RecordPerformance Test Interval DC CAL Output Test AC CAL Output Test Specification Equipment Required Critical Specification Model/Part NumberInput Resistance Page Voltage Measurement Accuracy Voltage Measurement Accuracy Test Div Offset Supply Upper Limit Lower Limit Procedure Zero-Input Offset Offset AccuracyDC Input Offset Div Offset Supply Min MaxBandwidth V1MHz Time Measurement Accuracy Menu Selection Setting Trigger Sensitivity RUN Repetitive Performance Test Record Paragraph No Test Results Yes No Page Table of Contents Page Calibration Protection Switch Calibration IntervalMaintaining System Integrity Instrument Warmup CalibrationSetting the PROTECT/UNPROTECT Switch to Unprotect Location of the PROTECT/UNPROTECT SwitchLoading the Default Calibration Factors Equipment Required for CalibrationCalibrate Delay of the Self Cal menu as follows Self Cal Menu CalibrationsCalibrate Vert Cal of the Self Cal menu as follows Calibrate the Time Null of the Self Cal menu as follows Calibrate the Logic Trigger of the Self Cal menu as followsCalibrate the Ext Trig Null of the Self Cal menu as follows Ext Trig Null Calibration ConnectionsSetting the PROTECT/UNPROTECT Switch to Protect Sealing the MainframeTable of Contents Page Replaceable Parts List AbbreviationsOrdering Information Exchange AssembliesDirect Mail Order System Reference Designator and Abbreviations Parts Identification Reference Designator HP Part Qty NumberDescription Mfr Code Mfr Part NumberPage Table of Contents Page Introduction Safety Requirements Module Block Diagram and The- ory of OperationMain Assembly Theory of Operation Attenuator Theory of OperationHP 16532A Oscilloscope Block Diagram Page Self-Test Access Procedure Self-TestsLoad Test System Screen Functional Tests Screen Touch the Functional Tests fieldA/D Test Run Screen Exit Test System Screen Self-Test Descriptions Data Memory TestTimebase Test TestTroubleshooting Troubleshooting AidsVerifying the System Clock Page Troubleshooting E E T Troubleshooting Sheet1 6 5 3 2 a Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Oscilloscope Module Removal and Replacement Oscilloscope Module Removal ProcedureOscilloscope Module Replacement Procedure 12. Endplate Overlap Bottom SequenceTrigger Cable Removal and Replacement Trigger Cable Removal ProcedureAttenuator Removal and Replacement Trigger Cable Replacement ProcedureAttenuator Removal Procedure Attenuator Replacement Procedure Warranty CertificationAssistance