HP 16532A - Performance Tests

Use the following table for steps 3 through 9.

V/Div

Offset

Supply

Upper Limit

Lower Limit

 

 

 

 

 

4 V/Div

-7.0 V

-14.0 V

-13.7 V

-14.3 V

1 V/Div

-1.75 V

-3.50 V

-3.43 V

-3.57 V

400 mV/Div

-700.0 mV

-1.40 V

-1.37 V

-1.43 V

40 mV/Div

-70.0 mV

-140.0 mV

-137.0 mV

-143.0 mV

40 mV/Div

70.0 mV

140.0 mV

143.0 mV

137.0 mV

400 mV/Div

700.0 mV

1.40 V

1.43 V

1.37 V

1 V/Div

1.75 V

3.50 V

3.57 V

3.43 V

4 V/Div

7.0 V

14.0 V

14.3 V

13.7 V

 

 

 

 

 

 

 

 

 

 

3.With the power supply disconnected from the channel input, set the Channel menu to the V/Div range and offset values shown in the first line of the table.

4.On the oscilloscope, select RUN - REPETITIVE and wait for approximately five seconds, then select STOP. Read the voltage displayed under the channel label on the left of the display. Enter this value on the performance test record in the "Zero Input" column.

5.Set the power supply to the first supply value listed in the table. Connect the power supply to the channel input as shown in figure 3-1.

6.On the oscilloscope, select RUN - REPETITIVE and wait for approximately five seconds, then select STOP. Read the voltage displayed under the channel label on the left of the display. Enter this value on the performance test record in the "Measured" column.

7.To obtain the test results, subtract the Zero Input voltage from the Measured voltage listed in the performance test record. For example: (Measured) - (Zero Input) = (Actual). When calculating the results, observe the plus or minus value of the voltage measurements. Record the results in the per- formance test record in the "Actual" column.

8.Repeat steps 3 through 8 for the second line of the table, then for the rest of the lines of the table for channel 1.

9.In the Waveform Selection menu, select Input channel 2. Delete channel 1 and insert channel 2. Re- peat steps 2 through 8 for channel 2, replacing channel 1 with channel 2 where applicable.

3-8

Image 34
Contents HP 16532A Safety Considerations Safety SymbolsContents Calibration ServiceSection Replaceable Parts List of Tables Page Table of Contents Page Introduction Safety RequirementsModule Covered by This Manual Specifications Accessories SuppliedAccessories Available Operating CharacteristicsWaveform Record Length 8000 points HP 16532A SpecificationsRise Time2 1.4 ns Vertical dc Gain Accuracy4 ± 1.5% of full scaleVertical at BNC HP 16532A Operating CharacteristicsTimebase Trigger Level Range TriggeringDigitizer Trigger ModesWaveform Display Display ModesMeasurement Aids Setup Aids Operating EnvironmentEquipment Critical Specifications Recommended Model USE Page Table of Contents Page Module Installation Power RequirementsInitial Inspection Safety Requirements Preparation for UseInstallation Considerations ProcedureEndplate Overlap Bottom Sequence Storage Tagging for ServiceOperating Environment PackagingTable of Contents Page Test Record Performance Test ProceduresPerformance Test Interval DC CAL Output Test AC CAL Output Test Equipment Required Critical Specification Model/Part Number SpecificationInput Resistance Page Voltage Measurement Accuracy Voltage Measurement Accuracy Test Div Offset Supply Upper Limit Lower Limit Procedure Zero-Input Offset Offset AccuracyDC Input Offset Div Offset Supply Min MaxBandwidth V1MHz Time Measurement Accuracy Menu Selection Setting Trigger Sensitivity RUN Repetitive Performance Test Record Paragraph No Test Results Yes No Page Table of Contents Page Calibration Interval Calibration Protection SwitchMaintaining System Integrity Instrument Warmup CalibrationSetting the PROTECT/UNPROTECT Switch to Unprotect Location of the PROTECT/UNPROTECT SwitchLoading the Default Calibration Factors Equipment Required for CalibrationSelf Cal Menu Calibrations Calibrate Delay of the Self Cal menu as followsCalibrate Vert Cal of the Self Cal menu as follows Calibrate the Time Null of the Self Cal menu as follows Calibrate the Logic Trigger of the Self Cal menu as followsCalibrate the Ext Trig Null of the Self Cal menu as follows Ext Trig Null Calibration ConnectionsSetting the PROTECT/UNPROTECT Switch to Protect Sealing the MainframeTable of Contents Page Ordering Information Replaceable Parts ListAbbreviations Exchange AssembliesDirect Mail Order System Reference Designator and Abbreviations Parts Identification Description Reference DesignatorHP Part Qty Number Mfr Code Mfr Part NumberPage Table of Contents Page Main Assembly Theory of Operation Introduction Safety RequirementsModule Block Diagram and The- ory of Operation Attenuator Theory of OperationHP 16532A Oscilloscope Block Diagram Page Self-Test Access Procedure Self-TestsLoad Test System Screen Functional Tests Screen Touch the Functional Tests fieldA/D Test Run Screen Exit Test System Screen Timebase Test Self-Test DescriptionsData Memory Test TestTroubleshooting Troubleshooting AidsVerifying the System Clock Page Troubleshooting Sheet Troubleshooting E E T1 6 5 3 2 a Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Oscilloscope Module Removal and Replacement Oscilloscope Module Removal ProcedureOscilloscope Module Replacement Procedure 12. Endplate Overlap Bottom SequenceTrigger Cable Removal and Replacement Trigger Cable Removal ProcedureTrigger Cable Replacement Procedure Attenuator Removal and ReplacementAttenuator Removal Procedure Attenuator Replacement Procedure Certification WarrantyAssistance