HP 16532A - Performance Tests
3-7. AC CAL OUTPUT Test
The AC CAL OUTPUT port on the rear panel is used for instrument calibration and probe compensation. Though calibrator accuracy is not specified in the performance specifications, it must be within limits in or- der to provide accurate calibration.
Test Limits:
1000 Hz ± 10%, 0.8 Vp_p ± 10%
Equipment Required:
The following equipment is required for the test.
|
| Recommended |
Equipment Required | Critical Specification | Model/Part Number |
|
|
|
Cable | 50 Ω BNC (m to m) | HP 10503A |
|
|
|
|
|
|
Procedure:
1.Use the BNC cable to connect the AC CAL OUTPUT to channel 1 input of the oscilloscope module.
2.On the oscilloscope module set the following parameters in the order given:
| Menu |
| Selection | Setting |
|
|
|
|
|
Calibration |
| Mode |
| Service Calibration |
|
| Procedure |
| Oscillator Output |
|
| Signal |
| Probe Compensation |
Channel |
| Coupling |
| 1 MΩ/DC |
|
| Probe |
| 1:1 |
|
|
|
|
|
|
|
|
|
|
3.In the Calibration menu select AUTOSCALE.
4.In the