HP 16532A - Performance Tests
3-13. TRIGGER SENSITIVITY
This test checks channel triggers for sensitivity at rated bandwidth.
Specification:
dc to 50 MHz : 0.063 x full scale (0.25 division)
50 to 250 MHz : 0.125 x full scale (0.5 division)
Equipment Required:
The following equipment is required for this test. Procedures are based on the model or part number rec- ommended.
|
|
| Recommended |
Equipment Required |
| Critical Specification | Model/Part Number |
|
|
| |
Signal Generator | 50 and 250 MHz, | HP 8656B Opt. 001 | |
Cable | 50 | Ω BNC | HP 10503B |
Adapter | 50 | Ω Type N (m) to BNC (f) | HP |
|
|
|
|
|
|
|
|
Procedure:
1.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. On the oscilloscope module set the following parameters in the order given:
| Menu |
| Selection |
| Setting |
|
|
|
|
|
|
Channel |
| Input |
| 1 |
|
|
| Probe |
| 1:1 | |
|
| v/Div |
| 400 mV | |
|
| Offset |
| 0 V | |
|
| Coupling |
| 50 | Ω / DC |
|
| s/Div |
| 20 ns | |
Trigger |
| Mode |
| Edge | |
|
| Source |
| Chan 1 | |
|
| Level |
| 0 V | |
Display |
| Mode |
| Average | |
|
| Average # |
| 8 |
|
|
| Grid |
| On |
|
|
| Markers |
| Off |
|
|
|
|
|
|
|
|
|
|
|
|
|
2. With an