HP 1GSa/s Digitizing Oscilloscope 16532A manual Ext Trig Null Calibration Connections

Page 55

HP 16532A - Calibration

Calibrate the Ext Trig Null of the Self Cal menu as follows:

1. Connect the cables, BNC adapter, and BNC tee adapter together as shown in figure 4-4.

Note

When you touch Start, the instrument will prompt you to connect the cables to the appropriate locations on the rear panel of the module.

Figure 4-4. Ext Trig Null Calibration Connections

2.Touch the Procedure field, then select Ext Trig Null from the pop-up menu.

3.Touch the Start field and follow the instructions on the display.

4.At the conclusion of the ext trig null calibration, remove the cables from the rear of the module.

Messages will be displayed as each calibration routine is completed to indicate calibration has passed or failed. The resulting calibration factors are automatically stored to nonvolatile RAM at the conclusion of each calibration routine.

This is the conclusion of the calibration procedures. The display should now look similar figure 4-5.

4-7

Image 55
Contents HP 16532A Safety Symbols Safety ConsiderationsContents Calibration ServiceSection Replaceable Parts List of Tables Page Table of Contents Page Introduction Safety RequirementsModule Covered by This Manual Operating Characteristics Accessories SuppliedAccessories Available SpecificationsVertical dc Gain Accuracy4 ± 1.5% of full scale HP 16532A SpecificationsRise Time2 1.4 ns Waveform Record Length 8000 pointsVertical at BNC HP 16532A Operating CharacteristicsTimebase Trigger Modes TriggeringDigitizer Trigger Level RangeDisplay Modes Waveform DisplayMeasurement Aids Operating Environment Setup AidsEquipment Critical Specifications Recommended Model USE Page Table of Contents Page Preparation for Use Power RequirementsInitial Inspection Safety Requirements Module InstallationProcedure Installation ConsiderationsEndplate Overlap Bottom Sequence Packaging Tagging for ServiceOperating Environment StorageTable of Contents Page Test Record Performance Test ProceduresPerformance Test Interval DC CAL Output Test AC CAL Output Test Equipment Required Critical Specification Model/Part Number SpecificationInput Resistance Page Voltage Measurement Accuracy Voltage Measurement Accuracy Test Div Offset Supply Upper Limit Lower Limit Offset Accuracy Procedure Zero-Input OffsetDiv Offset Supply Min Max DC Input OffsetBandwidth V1MHz Time Measurement Accuracy Menu Selection Setting Trigger Sensitivity RUN Repetitive Performance Test Record Paragraph No Test Results Yes No Page Table of Contents Page Calibration Interval Calibration Protection SwitchMaintaining System Integrity Calibration Instrument WarmupLocation of the PROTECT/UNPROTECT Switch Setting the PROTECT/UNPROTECT Switch to UnprotectEquipment Required for Calibration Loading the Default Calibration FactorsSelf Cal Menu Calibrations Calibrate Delay of the Self Cal menu as followsCalibrate Vert Cal of the Self Cal menu as follows Calibrate the Logic Trigger of the Self Cal menu as follows Calibrate the Time Null of the Self Cal menu as followsExt Trig Null Calibration Connections Calibrate the Ext Trig Null of the Self Cal menu as followsSealing the Mainframe Setting the PROTECT/UNPROTECT Switch to ProtectTable of Contents Page Exchange Assemblies Replaceable Parts ListAbbreviations Ordering InformationDirect Mail Order System Reference Designator and Abbreviations Parts Identification Mfr Code Mfr Part Number Reference DesignatorHP Part Qty Number DescriptionPage Table of Contents Page Attenuator Theory of Operation Introduction Safety RequirementsModule Block Diagram and The- ory of Operation Main Assembly Theory of OperationHP 16532A Oscilloscope Block Diagram Page Self-Tests Self-Test Access ProcedureLoad Test System Screen Touch the Functional Tests field Functional Tests ScreenA/D Test Run Screen Exit Test System Screen Test Self-Test DescriptionsData Memory Test Timebase TestTroubleshooting Aids TroubleshootingVerifying the System Clock Page Troubleshooting Sheet Troubleshooting E E T1 6 5 3 2 a Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Oscilloscope Module Removal Procedure Oscilloscope Module Removal and Replacement12. Endplate Overlap Bottom Sequence Oscilloscope Module Replacement ProcedureTrigger Cable Removal Procedure Trigger Cable Removal and ReplacementTrigger Cable Replacement Procedure Attenuator Removal and ReplacementAttenuator Removal Procedure Attenuator Replacement Procedure Certification WarrantyAssistance